8MM,MICRO BIPOLAR FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
Published December 3, 2014 · Recall initiated: November 3, 2014
Summary
FDA records medical device Class II recall Z-0431-2015, published December 3, 2014 and initiated November 3, 2014. Recalling firm: Intuitive Surgical, Inc.; Status: Terminated; Reported quantity: 509 Units, total all affected units; Lot: Part number: 470171-05; All 8mm da Vinci Xi (IS4000) Surgical System instruments manufactured with white colored housings. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Intuitive Surgical, Inc.
- Distributor
- US Distribution -- FL, TX, NY, NV, OR, IN, UT, and MA.
- Lot
- Part number: 470171-05; All 8mm da Vinci Xi (IS4000) Surgical System instruments manufactured with white colored housings.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15July 22, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/AJune 10, 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06May 13, 2026
- FDA8MM,LONG TIP FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.December 3, 2014
- FDA8MM,PROGRASP FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.December 3, 2014
Questions about this recall
Was 8MM,MICRO BIPOLAR FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments. recalled?
Yes. FDA published recall Z-0431-2015 on December 3, 2014. Reported hazard: Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
What should I do if I have this product?
Voluntary: Firm initiated · Letter