8MM, CURVED BIPOLAR DISSECTOR; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Published December 3, 2014 · Recall initiated: November 3, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Intuitive Surgical, Inc.
Distributor
US Distribution -- FL, TX, NY, NV, OR, IN, UT, and MA.
Lot
Part number; 470344-04; All 8mm da Vinci Xi (IS4000) Surgical System instruments manufactured with white colored housings.
Risk classification
Class II
Area (FDA)
device
Country
United States