Neptune® 2 Rover - ULTRA ~120V ~ 60 Hz 12A REF 0702-001-000, Rx, Caution, consult accompanying documents Consult instructions for use, Max Vacuum; 21.0 in-Hg, 530 mm-Hg, The Neptune Waste Management System consists of a mobile rover unit used to suction and collect fluid waste, small debris and electrocautery smoke from a surgical site. Mobility allows the rover unit to be relocated to a waste disposal area where the rovers collection canisters can be emptied, via the docking station. Intended to be used in the Operating Room, Surgical Centers and Doctors Offices to collect and dispose of surgical fluid waste as well as collect smoke from electrocautery or laser devices.
Published December 12, 2012 · Recall initiated: May 24, 2011
Summary
FDA records medical device Class II recall Z-0454-2013, published December 12, 2012 and initiated May 24, 2011. Recalling firm: Stryker Instruments Div. of Stryker Corporation; Status: Terminated; Reported quantity: 5865; Lot: Part 0702-001-000 (120v) Lot 0733700353 through 1036100423. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential for the casters to bind compromising the Rover's ability to roll. This can result in additional strain on the operator, due to multiple exposures resulting in back pain.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Instruments Div. of Stryker Corporation
- Distributor
- Worldwide Distribution-USA (nationwide) and Pacific countries, and the countries of EEMEA, Canada, Latin America, and Japan.
- Lot
- Part 0702-001-000 (120v) Lot 0733700353 through 1036100423
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting,July 27, 2022
- FDANeptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The Neptune E-SEP is anMay 6, 2020
- FDAColor Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder), 1PRT(Single Port) QuickDecember 11, 2019
- FDAC2 Nerve Monitor, 8 Channel, 5140-508-280, UDI 07613327277319 Product Usage: The C2 Nerve Monitor System isOctober 9, 2019
- FDAOmniGuide BP-ROBOTIC ( BeamPath Robotic), Single Use Cat. # BP-ROBOTIC The OmniGuide Fiber is indicated forDecember 12, 2012
- FDANeptune® 2 Rover - ULTRA ~230V ~ 50 Hz 12A REF 0702-002-000, Rx Caution, consult accompanying documentsDecember 12, 2012
Questions about this recall
Was Neptune® 2 Rover - ULTRA ~120V ~ 60 Hz 12A REF 0702-001-000, Rx, Caution, consult accompanying documents Consult instructions for use, Max Vacuum; 21.0 in-Hg, 530 mm-Hg, The Neptune Waste Management System consists of a mobile rover unit used to suction and collect fluid waste, small debris and electrocautery smoke from a surgical site. Mobility allows the rover unit to be relocated to a waste disposal area where the rovers collection canisters can be emptied, via the docking station. Intended to be used in the Operating Room, Surgical Centers and Doctors Offices to collect and dispose of surgical fluid waste as well as collect smoke from electrocautery or laser devices. recalled?
Yes. FDA published recall Z-0454-2013 on December 12, 2012. Reported hazard: Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential for the casters to bind compromising the Rover's ability to roll. This can result in additional strain on the operator, due to multiple exposures resulting in back pain.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit