Neptune¿ 2 Rover - ULTRA ~230V ~ 50 Hz 12A REF 0702-002-000, Rx Caution, consult accompanying documents Consult instructions for use, Max Vacuum; 21.0 in-Hg, 530 mm-Hg, The Neptune Waste Management System consists of a mobile rover unit used to suction and collect fluid waste, small debris and electrocautery smoke from a surgical site. Mobility allows the rover unit to be relocated to a waste disposal area where the rovers collection canisters can be emptied, via the docking station. Intended to be used in the Operating Room, Surgical Centers and Doctors Offices to collect and dispose of surgical fluid waste as well as collect smoke from electrocautery or laser devices.

Published December 12, 2012 · Recall initiated: May 24, 2011

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential for the casters to bind compromising the Rover's ability to roll. This can result in additional strain on the operator, due to multiple exposures resulting in back pain.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Stryker Instruments Div. of Stryker Corporation
Distributor
Worldwide Distribution-USA (nationwide) and Pacific countries, and the countries of EEMEA, Canada, Latin America, and Japan.
Lot
Part 0702-002-000 (230v) Lot 0733700353 through 1036100423
Risk classification
Class II
Area (FDA)
device
Country
United States