iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological
Published December 18, 2013 · Recall initiated: August 20, 2013
Summary
FDA records medical device Class II recall Z-0456-2014, published December 18, 2013 and initiated August 20, 2013. Recalling firm: Iba Dosimetry Gmbh; Status: Terminated; Reported quantity: 33 units; Lot: SW Version 3.0. Catalong Number CS10-100. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Iba Dosimetry Gmbh
- Distributor
- Worldwide Distribution - US Distribution including the states of CT, MA, OH, CA, SC, NY and OK, and the countries of International distribution (from Germany) Poland, Germany, GB, The Netherlands, Spain, Turkey, Canada, Mexico, China, Hong Kong and Singapore.
- Lot
- SW Version 3.0. Catalong Number CS10-100
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDACOMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiologicalNovember 12, 2014
- FDACutera brand truSculpt infrared System (optional), Massager, Vacuum, radio frequency induced heat, Product isDecember 18, 2013
- FDAProduct is labeled in part: LIST NO. 43449; RECEPTAL CANISTER; DO NOT FILL 900 ABOVE THIS LINE; Product ofDecember 18, 2013
- FDAVitacost Super Daily Enzymes 500 mg caps, product is packaged in a plastic bottle with 240 capsules perDecember 18, 2013
- FDAEthicon Stratafix Symmetric PDS Plus Knotless Tissue Control Device Ethicon LLC Guaynabo, Puerto Rico 00969.December 18, 2013
- FDADigestive Enzyme Complex/Enzyme Formula Caps, Quantity:13M, Product #: IB286, product is packaged in bulkDecember 18, 2013
Questions about this recall
Was iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological recalled?
Yes. FDA published recall Z-0456-2014 on December 18, 2013. Reported hazard: The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit