iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological

Published December 18, 2013 · Recall initiated: August 20, 2013

Summary

FDA records medical device Class II recall Z-0456-2014, published December 18, 2013 and initiated August 20, 2013. Recalling firm: Iba Dosimetry Gmbh; Status: Terminated; Reported quantity: 33 units; Lot: SW Version 3.0. Catalong Number CS10-100. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Iba Dosimetry Gmbh
Distributor
Worldwide Distribution - US Distribution including the states of CT, MA, OH, CA, SC, NY and OK, and the countries of International distribution (from Germany) Poland, Germany, GB, The Netherlands, Spain, Turkey, Canada, Mexico, China, Hong Kong and Singapore.
Lot
SW Version 3.0. Catalong Number CS10-100
Risk classification
Class II
Area (FDA)
device
Country
Germany

Related recalls

Questions about this recall

Was iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological recalled?

Yes. FDA published recall Z-0456-2014 on December 18, 2013. Reported hazard: The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit