Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully automated system for clinical chemistry analysis.
Published December 12, 2012 · Recall initiated: November 1, 2012
Summary
FDA records medical device Class II recall Z-0473-2013, published December 12, 2012 and initiated November 1, 2012. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: 78 analyzers; Lot: Catalog numbers affected: Alkaline Phosphatase acc. to IFCC Gen.2 05166888190; Alanine Aminotransferase acc. to IFCC without pyridoxal phosphate (P5P) activation 05850797190; Bicarbonate Liquid 05446376190; Creatinine Jaff® Gen.2 05168597190; C-Reactive Protein Gen.3 05172373190; and Urea/BUN 05171873190. Serial Number(s): 0917-10, 1020-04, 1020-05, 1021-03, 1021-04, 1021-07, 1021-08, 1023-06, 1023-09, 1023-10, 1024-03, 1024-04, 1024-05, 1024-09, 1025-02, 1025-05, 1025-07, 1025-09, 1026-01, 1026-02, 1027-09, 1028-01, 1028-07, 1030-09, 1031-05, 1031-09, 1032-09, 1032-10, 1033-04, 1033-05, 1107-06, 1109-02, 1109-03, 1111-07, 1111-08, 1112-02, 1112-04, 1115-06, 1115-07, 1115-08, 1115-10, 1117-04, 1117-07, 1118-02, 1126-02, 1126-03, 1126-04, 1127-01, 1127-02, 1127-06, 1127-07, 1138-03, 1138-04, 1139-04, 1139-05, 1139-06, 1140-02, 1140-03, 1140-04, 1140-05, 1140-10, 1141-08, 1141-09, 1142-02, 1142-03, 1142-04, 1143-01, 1143-02, 1143-03, 1143-04, 1143-05, 1143-06, 1144-05, 1146-04, 1146-05, 1146-06, 1146-07, and 1146-08. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Roche Diagnostics Operations, Inc.
- Distributor
- Nationwide Distribution, including the states of AL, AZ, CA, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, NJ, NY, OH, VA, WA, and WI.
- Lot
- Catalog numbers affected: Alkaline Phosphatase acc. to IFCC Gen.2 05166888190; Alanine Aminotransferase acc. to IFCC without pyridoxal phosphate (P5P) activation 05850797190; Bicarbonate Liquid 05446376190; Creatinine Jaff® Gen.2 05168597190; C-Reactive Protein Gen.3 05172373190; and Urea/BUN 05171873190. Serial Number(s): 0917-10, 1020-04, 1020-05, 1021-03, 1021-04, 1021-07, 1021-08, 1023-06, 1023-09, 1023-10, 1024-03, 1024-04, 1024-05, 1024-09, 1025-02, 1025-05, 1025-07, 1025-09, 1026-01, 1026-02, 1027-09, 1028-01, 1028-07, 1030-09, 1031-05, 1031-09, 1032-09, 1032-10, 1033-04, 1033-05, 1107-06, 1109-02, 1109-03, 1111-07, 1111-08, 1112-02, 1112-04, 1115-06, 1115-07, 1115-08, 1115-10, 1117-04, 1117-07, 1118-02, 1126-02, 1126-03, 1126-04, 1127-01, 1127-02, 1127-06, 1127-07, 1138-03, 1138-04, 1139-04, 1139-05, 1139-06, 1140-02, 1140-03, 1140-04, 1140-05, 1140-10, 1141-08, 1141-09, 1142-02, 1142-03, 1142-04, 1143-01, 1143-02, 1143-03, 1143-04, 1143-05, 1143-06, 1144-05, 1146-04, 1146-05, 1146-06, 1146-07, and 1146-08.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAcobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, inNovember 27, 2024
- FDACreatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitativeOctober 2, 2024
- FDAAlbumin Gen.2, Catalog 05166861190October 21, 2020
- FDAElecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virusDecember 18, 2019
- FDAOSS Reamer Sleeve REF132-472662 KNEE REAMER SLEEVE 13.0 thru 20.0 STAINLESS STEEL The OSS Knee Reamer SleevesDecember 12, 2012
- FDABio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant) Product Usage: The OSS KneeDecember 12, 2012
Questions about this recall
Was Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully automated system for clinical chemistry analysis. recalled?
Yes. FDA published recall Z-0473-2013 on December 12, 2012. Reported hazard: On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C
What should I do if I have this product?
Voluntary: Firm initiated · Letter