Back to search
FDAFDA · Food and drugs·Recall number Z-0474-2013

Bio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant) Product Usage: The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.

Published December 12, 2012 · Recall initiated: November 1, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly bearing. Increased wear could potentially lead to a shorter implant life.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Biomet, Inc.
Distributor
USA Nationwide Distribution in the state of WI.
Lot
Catalog Number: 113740 and Lot Number Identification: 779270
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card