Back to search
FDAFDA · Food and drugs·Recall number Z-0473-2025

Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4704

Published November 27, 2024 · Recall initiated: October 4, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Hobbs Medical, Inc.
Distributor
Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.
Lot
Catalog Number: 4704; UDI-DI: M84947040; Lot Numbers: H10-23-106, H11-23-059.
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.