Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4722
Published November 27, 2024 · Recall initiated: October 4, 2024
Summary
FDA records medical device Class II recall Z-0475-2025, published November 27, 2024 and initiated October 4, 2024. Recalling firm: Hobbs Medical, Inc.; Status: Ongoing; Reported quantity: 198 units; Lot: Catalog Number: 4722; UDI-DI: M84947220; Lot Numbers: H08-23-265, H09-23-113, H10-23-223, H11-23-058. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hobbs Medical, Inc.
- Distributor
- Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.
- Lot
- Catalog Number: 4722; UDI-DI: M84947220; Lot Numbers: H08-23-265, H09-23-113, H10-23-223, H11-23-058.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPosi-Stop Injection Needle, 23 Gauge, Catalog Number: 4721November 27, 2024
- FDAPosi-Stop Injection Needle, 25 Gauge, Catalog Number: 4718November 27, 2024
- FDAPosi-Stop Injection Needle, 23 Gauge, Catalog Number: 4716November 27, 2024
- FDAPosi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723November 27, 2024
- FDAPosi-Stop Injection Needle, 25 Gauge, Catalog Number: 4704November 27, 2024
- FDAJuggerLoc/ToggleLoc 2.9mm Drill Guide, Model number 110027357November 27, 2024
Questions about this recall
Was Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4722 recalled?
Yes. FDA published recall Z-0475-2025 on November 27, 2024. Reported hazard: Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
What should I do if I have this product?
Voluntary: Firm initiated · Letter