Back to search
FDAFDA · Food and drugs·Recall number Z-0490-2016

BZO Strip Dip Card (300ng/ml)*, Item No. 100170

Published December 30, 2015 · Recall initiated: November 16, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Ameditech Inc
Distributor
Nationwide Distribution.
Lot
141233 141669 141790 141820 141958 142082 142531 142532 142835 143020 143471 143553 152021 152432 152546 153094 153538 152104
Risk classification
Class III
Area (FDA)
device
Country
United States