FDAFDA · Food and drugs·Recall number Z-0494-2016
DrugSmart 11 Test Cup, Item No. 61127D
Published December 30, 2015 · Recall initiated: November 16, 2015
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Ameditech Inc
- Distributor
- Nationwide Distribution.
- Lot
- 144196 144375 150096 151980 152737
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.