Back to search
FDAFDA · Food and drugs·Recall number Z-0492-2021

Geistlich Bio-Oss Pen 1-2mm 0.5g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20126

Published December 2, 2020 · Recall initiated: October 19, 2020

Summary

FDA records medical device Class II recall Z-0492-2021, published December 2, 2020 and initiated October 19, 2020. Recalling firm: Geistlich Pharma North America, Inc.; Status: Terminated; Reported quantity: 537 units; Lot: Lots: 81901651,82000466. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Geistlich Pharma North America, Inc.
Distributor
Worldwide distribution - US Nationwide distribution and the country of Japan.
Lot
Lots: 81901651,82000466
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Geistlich Bio-Oss Pen 1-2mm 0.5g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20126 recalled?

Yes. FDA published recall Z-0492-2021 on December 2, 2020. Reported hazard: Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane

What should I do if I have this product?

Voluntary: Firm initiated · Letter