Geistlich Bio-Oss Pen 1-2mm 0.5g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20126
Published December 2, 2020 · Recall initiated: October 19, 2020
Summary
FDA records medical device Class II recall Z-0492-2021, published December 2, 2020 and initiated October 19, 2020. Recalling firm: Geistlich Pharma North America, Inc.; Status: Terminated; Reported quantity: 537 units; Lot: Lots: 81901651,82000466. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Geistlich Pharma North America, Inc.
- Distributor
- Worldwide distribution - US Nationwide distribution and the country of Japan.
- Lot
- Lots: 81901651,82000466
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAChlorhexidine Gluconate Oral Rinse USP, 0.12%, Alcohol Free, packaged in 15 mL unit does cups (barcodeDecember 9, 2020
- FDABox Label: ION Fully Articulating Catheter, REF 490105, VER 08 - Product Usage: used to navigate the Ion FullyDecember 2, 2020
- FDACardioblate CryoFlex, Surgical Ablation Probe, Model Number 60SF7, PROBE 60SF7 CRYOFLEX 7CM 26L - is a singleDecember 2, 2020
- FDAAB PowerCel 170 (velvet black), REF: CI-5517-150, UDI: (01) 07630016833113 used to power the followingDecember 2, 2020
- FDACardioblate CryoFlex, Surgical Ablation Probe, Model Number 60CM1, CLAMP 60CM1 CRYOFLEX PROBE 26L - is aDecember 2, 2020
- FDAViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency RangeDecember 2, 2020
Questions about this recall
Was Geistlich Bio-Oss Pen 1-2mm 0.5g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20126 recalled?
Yes. FDA published recall Z-0492-2021 on December 2, 2020. Reported hazard: Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane
What should I do if I have this product?
Voluntary: Firm initiated · Letter