Back to search
FDAFDA · Food and drugs·Recall number Z-0498-2025

Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Published November 27, 2024 · Recall initiated: September 18, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors evaluation of the GIF-1TH190.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
Domestic: US Nationwide Distribution.
Lot
Model: GIF-1TH190, UDI-DI: 04953170343360, Serial #: Less than 2446103.
Risk classification
Class II
Area (FDA)
device
Country
United States