FDAFDA · Food and drugs·Recall number Z-0498-2025
Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.
Published November 27, 2024 · Recall initiated: September 18, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors evaluation of the GIF-1TH190.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- Domestic: US Nationwide Distribution.
- Lot
- Model: GIF-1TH190, UDI-DI: 04953170343360, Serial #: Less than 2446103.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States