FDAFDA · Food and drugs·Recall number Z-0507-2016

Immutest 10 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-10M5

Published December 30, 2015 · Recall initiated: November 16, 2015

Summary

FDA records medical device Class III recall Z-0507-2016, published December 30, 2015 and initiated November 16, 2015. Recalling firm: Ameditech Inc; Status: Terminated; Reported quantity: 320 kits; Lot: 143846 144380 150312 150505 153177 153488. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ameditech Inc
Distributor
Nationwide Distribution.
Lot
143846 144380 150312 150505 153177 153488
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Immutest 10 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-10M5 recalled?

Yes. FDA published recall Z-0507-2016 on December 30, 2015. Reported hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

What should I do if I have this product?

Voluntary: Firm initiated · Letter