Back to search
FDAFDA · Food and drugs·Recall number Z-0509-2016

Immutest 5 Pnl Drug Cup w/ Adult -waived, Item No. IMCA-5AB-W

Published December 30, 2015 · Recall initiated: November 16, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Ameditech Inc
Distributor
Nationwide Distribution.
Lot
141758 141921 142657 143178 143581 144563 151337 152240 152782 153154 153401
Risk classification
Class III
Area (FDA)
device
Country
United States