Back to search
FDAFDA · Food and drugs·Recall number Z-0518-2024

Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix Knotless anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder.

Published December 20, 2023 · Recall initiated: October 25, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Riverpoint Medical, LLC
Distributor
U.S. Nationwide distribution in the state of CA.
Lot
Product Code: 3911-7 14-520 UDI-DI Code: 00840277402172 Lot Number: 23080433 Product Code: 3911-7 14-521 UDI-DI Code: 00840277402189 Lot Number: 23081617
Risk classification