Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix Knotless anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder.
Published December 20, 2023 · Recall initiated: October 25, 2023
Summary
FDA records medical device Class II recall Z-0518-2024, published December 20, 2023 and initiated October 25, 2023. Recalling firm: Riverpoint Medical, LLC; Status: Ongoing; Reported quantity: 999 devices; Lot: Product Code: 3911-7 14-520 UDI-DI Code: 00840277402172 Lot Number: 23080433 Product Code: 3911-7 14-521 UDI-DI Code: 00840277402189 Lot Number: 23081617. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Riverpoint Medical, LLC
- Distributor
- U.S. Nationwide distribution in the state of CA.
- Lot
- Product Code: 3911-7 14-520 UDI-DI Code: 00840277402172 Lot Number: 23080433 Product Code: 3911-7 14-521 UDI-DI Code: 00840277402189 Lot Number: 23081617
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPanorama 1.0T HFO, Magnetic Resonance (MR) System; Model Numbers 781250 and 781350. Philips Magnetic ResonanceDecember 20, 2023
- FDASTERILE PVP SOLUTION, REF DYNDA1907December 20, 2023
- FDAVelosorb Fast Braided Absorbable Suture: Covidien Velosorb 5/0 Undyed 30" KV-11, TaperCutting, Product Number:October 7, 2020
- FDAVelosorb Fast Braided Absorbable Suture: Covidien Velosorb 4/0 Undyed 30" P-12 Cutting SV426, UDI:October 7, 2020
- FDAVelosorb Fast Braided Absorbable Suture: Covidien Velosorb 4/0 Undyed 30" CV-25 Taper, Product Number: CV3100,October 7, 2020
- FDAVelosorb Fast Braided Absorbable Suture: Covidien Velosorb 0 Undyed 36" GS-21 Taper CV946, UDI: 20884521151298October 7, 2020
Questions about this recall
Was Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix Knotless anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder. recalled?
Yes. FDA published recall Z-0518-2024 on December 20, 2023. Reported hazard: Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter