FDAFDA · Food and drugs·Recall number Z-0535-2025

VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740

Published December 4, 2024 · Recall initiated: October 11, 2024

Summary

FDA records medical device Class II recall Z-0535-2025, published December 4, 2024 and initiated October 11, 2024. Recalling firm: Ortho-Clinical Diagnostics, Inc.; Status: Ongoing; Reported quantity: 7717 units; Lot: UDI/GTIN 10758750000296 Coating 1221, expiration date 10/01/2024 Coating 1222, expiration date 01/01/2025. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ortho-Clinical Diagnostics, Inc.
Distributor
Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS, MN, MT, NY, PA, UT and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 0B8, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
Lot
UDI/GTIN 10758750000296 Coating 1221, expiration date 10/01/2024 Coating 1222, expiration date 01/01/2025
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740 recalled?

Yes. FDA published recall Z-0535-2025 on December 4, 2024. Reported hazard: May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.

What should I do if I have this product?

Voluntary: Firm initiated