FDAFDA · Food and drugs·Recall number Z-2580-2024
VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2
Published August 21, 2024 · Recall initiated: July 3, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition codes which lead to a delay of results.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Sweden, United Kingdom.
- Lot
- Model No 1513266; UDI/GTIN# 110758750009237; Lots within expiry: 3380, 3390, 3400, 3410, 3422, 3430, 3440, 3450, 3461, 3471, 3480.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States