VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2
Published August 21, 2024 · Recall initiated: July 3, 2024
Summary
FDA records medical device Class II recall Z-2580-2024, published August 21, 2024 and initiated July 3, 2024. Recalling firm: Ortho-Clinical Diagnostics, Inc.; Status: Ongoing; Reported quantity: 18,467 units; Lot: Model No 1513266; UDI/GTIN# 110758750009237; Lots within expiry: 3380, 3390, 3400, 3410, 3422, 3430, 3440, 3450, 3461, 3471, 3480. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition codes which lead to a delay of results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Sweden, United Kingdom.
- Lot
- Model No 1513266; UDI/GTIN# 110758750009237; Lots within expiry: 3380, 3390, 3400, 3410, 3422, 3430, 3440, 3450, 3461, 3471, 3480.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP)December 4, 2024
- FDAThe VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and aAugust 28, 2024
- FDAVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd.August 21, 2024
- FDAHeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS)August 21, 2024
- FDAVITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 ChemistryJuly 10, 2024
- FDAVITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROSJuly 10, 2024
Questions about this recall
Was VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2 recalled?
Yes. FDA published recall Z-2580-2024 on August 21, 2024. Reported hazard: It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition codes which lead to a delay of results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter