FDAFDA · Food and drugs·Recall number Z-0553-2025
Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
Published December 4, 2024 · Recall initiated: October 9, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Shimadzu Corp. Analytical & Measuring Instruments Div.
- Distributor
- Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.
- Lot
- Model/UDI-DI/Serial Number: SIL-20ACXR CL/04540217420131/L20995250001, L20995250002, L20995250003, L20995250004, L20995250005, L20995250006, L20995250007, L20995250008, L20995350010, L20995350011, L20995350012, L20995350013, L20995350014, L20995350015, L20995450016, L20995450017, L20995450018, L20995450019, L20995450020, L20995450021, L20995450022, L20995450023, L20995450024, L20995450025, L20995450026, L20995550027, L20995550028, L20995550029, L20995550030, L20995650031, L20995650032, L20995650033, L20995650034, L20995650035, L20995650036, L20995650037, L20995650038, L20995650039, L20995650040, L20995650041, L20995650042, L20995650043, L20995650044, L20995650045, L20995650046, L20995560001, L20995660002, L20995660003, L20995660004, L20995660005, L20995660006, L20995660007, L20995660008, L20995660009, L20995660010, L20995660011, L20995660012, L20995660013, L20995660014, L20995660015, L20995660016, L20995660017, L20995660018, L20995660019, L20995660020, L20995660021, L20995660022, L20995660023, L20995660024, L20995660025, L20995660026, L20995660027, L20995660028, L20995660029, L20995660030, L20995660031, L20995660032, L20995660033, L20995660034, L20995660035, L20995660036, L20995660037, L20995660038, L20995660039, L20995660040, L20995660041, L20995660042, L20995660043, L20995660044, L20995660045, L20995660046, L20995660047, L20995660048, L20995660049, L20995660050, L20995660051, L20995660052, L20995660053, L20995660054, L20995660055, L20995660056, L20995660057, L20995250009, L20995270001, L20995270002, L20995270003, L20995270004, L20995270005, L20995370006, L20995370007, L20995370008, L20995370009, L20995370010, L20995370011, L20995370012, L20995370013, L20995370014, L20995370015, L20995370016, L20995370017, L20995370018, L20995370019, L20995370020, L20995370021, L20995370022, L20995370023, L20995370024, L20995370025, L20995470028, L20995770032. SIL-20ACHT CL/04540217420124/L20985350001, L20985450002, L20985750003, L20985750004. SIL-30AC CL/04540217420155/L21005250001, L21005250002, L21005250003, L21005250004, L21005250005, L21005250006, L21005250007, L21005350008, L21005350009, L21005350010, L21005350011, L21005350012, L21005350013, L21005350014, L21005350015, L21005350016, L21005350017, L21005350018, L21005350019, L21005350020, L21005350021, L21005350022, L21005350023, L21005350024, L21005450025, L21005550026, L21005550027, L21005650028, L21005650029, L21005750030, L21005750031, L21005570001, L21005570002, L21005670003, L21005670004, L21005670005, L21005670006, L21005670007, L21005670008, L21005770009, L21005770010, L21005770011, L21005770012, L21005770013, L21005770014.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Japan
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.