Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Published December 4, 2024 · Recall initiated: October 9, 2024
Summary
FDA records medical device Class II recall Z-0554-2025, published December 4, 2024 and initiated October 9, 2024. Recalling firm: Shimadzu Corp. Analytical & Measuring Instruments Div.; Status: Ongoing; Reported quantity: 48; Lot: Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Shimadzu Corp. Analytical & Measuring Instruments Div.
- Distributor
- Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
- Lot
- Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Japan
Related recalls
- FDAAutosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CLDecember 4, 2024
- FDATempus LS-Manual Defibrillator Model 00-3020 The Tempus LS-Manual is a defibrillator that can deliver a shockDecember 4, 2024
- FDAMerlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the MerlinOnDemand"December 4, 2024
- FDAENDO KIT, Medline Kit SKU DYKE1577A, Component Number 57469, 10 per case; Medline Convenience Kits containingDecember 4, 2024
- FDAMerlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data fromDecember 4, 2024
- FDABaxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4482; use in Peritoneal DialysisDecember 4, 2024
Questions about this recall
Was Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL recalled?
Yes. FDA published recall Z-0554-2025 on December 4, 2024. Reported hazard: There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter