Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
Published January 1, 2014 · Recall initiated: November 6, 2013
Summary
FDA records medical device Class II recall Z-0571-2014, published January 1, 2014 and initiated November 6, 2013. Recalling firm: Data Innovations, Inc.; Status: Terminated; Reported quantity: 24 systems; Lot: Software alll versions prior to: v7.00.0006 and v8.00. Includes: v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Data Innovations, Inc.
- Distributor
- Nationwide Canada and Britain
- Lot
- Software alll versions prior to: v7.00.0006 and v8.00. Includes: v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for theJanuary 1, 2014
- FDACaptia Syphilis-G Test Kit Product Usage: This is an enzyme immunoassay for the qualitative detection of IgGJanuary 1, 2014
- FDAIMMULITE/IMMULITE 1000 Systems FOL Folic Acid, REF/Catalog Number LKFO1 (100 tests), Siemens Material NumberJanuary 1, 2014
- FDAIndus Invue Screws: IM71058-XX: 04.2mm, SelfTapping, Tapered. Used to secure the Invue Anterior Cervical PlateJanuary 1, 2014
- FDAZiehm Vision RFD C-Arm Interventional Fluoroscopic X-Ray System. Mobile Fluoroscopic ImagingJanuary 1, 2014
- FDAIndus Invue Screws: lM71059-XX: Indus Screw 04.2mm SelfDrilling. Used to secure the Invue Anterior CervicalJanuary 1, 2014
Questions about this recall
Was Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004 recalled?
Yes. FDA published recall Z-0571-2014 on January 1, 2014. Reported hazard: When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail