Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp - Extra Short, 5C4483
Published February 16, 2022 · Recall initiated: January 7, 2022
Summary
FDA records medical device Class II recall Z-0572-2022, published February 16, 2022 and initiated January 7, 2022. Recalling firm: Baxter Healthcare Corporation; Status: Ongoing; Reported quantity: 344,190 units; Lot: All lot numbers within expiry. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- Worldwide
- Lot
- All lot numbers within expiry
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATruSystem 7500 surgical table, Model/Catalog Number: 4091000November 11, 2026
- FDATruSystem 7000 U Operating Table (dV), Product Code 1723633November 11, 2026
- FDATruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087November 11, 2026
- FDAOperating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788November 11, 2026
- FDABaxter MiniCap Extended Life PD Transfer Set (Easy-Lock), 5C4449February 16, 2022
- FDABaxter MiniCap Extended Life PD Transfer Set with Twist Clamp, REF R5C4482EFebruary 16, 2022
Questions about this recall
Was Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp - Extra Short, 5C4483 recalled?
Yes. FDA published recall Z-0572-2022 on February 16, 2022. Reported hazard: Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
What should I do if I have this product?
Voluntary: Firm initiated · Letter