FDAFDA · Food and drugs·Recall number Z-0572-2022
Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp - Extra Short, 5C4483
Published February 16, 2022 · Recall initiated: January 7, 2022
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- Worldwide
- Lot
- All lot numbers within expiry
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?