OrthoPediatrics ACL Reconstruction System is a pediatric and small stature adult ACL reconstruction system containing Sleeve Installation Tool Set , Part 01-1008-0040/0047, 6mm; 6.5mm, 7mm, 7.5mm, 8mm, 8.5mm, 9mm, and 10mm. Used with ACL reconstruction system associated surgery.
Published January 1, 2014 · Recall initiated: November 7, 2013
Summary
FDA records medical device Class II recall Z-0579-2014, published January 1, 2014 and initiated November 7, 2013. Recalling firm: OrthoPediatrics Corp; Status: Terminated; Reported quantity: 40; Lot: Part 01-1008-0040/0047 Lot 132762, F132763, F132764, F132765, F132766, F132768, F132769, F132770. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Part # 01-1008-0040/0047, ACL Sleeve Installation/ Insertion Tool, shipped November 4, 2013 - November 6, 2013 may be etched with the wrong size.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- OrthoPediatrics Corp
- Distributor
- Nationwide Distribution including the states of CA, KY, OH, RI, and TN.
- Lot
- Part 01-1008-0040/0047 Lot 132762, F132763, F132764, F132765, F132766, F132768, F132769, F132770
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOrthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended forAugust 5, 2020
- FDAOrthoPediatrics ACL Reconstruction System ShieldLoc 7mm implant component model number 10-1008-4070 - ProductJuly 15, 2020
- FDAModel No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform. ProductNovember 6, 2019
- FDAResponse BandLoc 5.5/6.0 Tensioner Instrument The Tensioner Instrument (01-1102-1000) is utilized to allow theJune 14, 2017
- FDAProPlan CMF Guide(s) Model(s) REF SD900.008 Description: Single Use Only : Patient-Specific Guide,January 1, 2014
- FDAsteroEOS Workstation with software versions: 1.3, 1.4 and 1.5 used in processing 3D X-ray images ProductJanuary 1, 2014
Questions about this recall
Was OrthoPediatrics ACL Reconstruction System is a pediatric and small stature adult ACL reconstruction system containing Sleeve Installation Tool Set , Part 01-1008-0040/0047, 6mm; 6.5mm, 7mm, 7.5mm, 8mm, 8.5mm, 9mm, and 10mm. Used with ACL reconstruction system associated surgery. recalled?
Yes. FDA published recall Z-0579-2014 on January 1, 2014. Reported hazard: Part # 01-1008-0040/0047, ACL Sleeve Installation/ Insertion Tool, shipped November 4, 2013 - November 6, 2013 may be etched with the wrong size.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit