Remote Patient Monitoring System, MDP 2000 Series, Part Numbers: MDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected Examples of areas where the MDP200 is used include ICU, CCU, PACU, Emergency, Telemetry Step-down and other areas where patient's physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values which as heart rate and systolic, mean and diastolic pressure. This device acts as the Central Monitor Station, it permits from one (1) to sixteen (16) bedside monitor(s) to be connected simultaneously.
Published January 13, 2016 · Recall initiated: November 8, 2015
Summary
FDA records medical device Class II recall Z-0582-2016, published January 13, 2016 and initiated November 8, 2015. Recalling firm: Vidco, Inc.; Status: Terminated; Reported quantity: 59 units; Lot: Serial numbers: 6533-6537, 6561, 6566-6581, 6539, 6540, 6558-6560, 6562, 6508-6528, 6563-6565, 6529, 6531, 6532, 6541, 6551, 6556, and 6557. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Testing at customer site showed unit Remote Patient Monitoring System MDP2040-0100 in a continuous trap condition, not allowing system to reset and reboot. Two customers recently complained of the system freezing and it could only be restarted if the user re-applied power.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Vidco, Inc.
- Distributor
- US Nationwide distribution in the states of AZ, CA, MD, NM, NJ, and OH.
- Lot
- Serial numbers: 6533-6537, 6561, 6566-6581, 6539, 6540, 6558-6560, 6562, 6508-6528, 6563-6565, 6529, 6531, 6532, 6541, 6551, 6556, and 6557.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABRIELLE ORTHO CONTENDER POST-OP-KNEE BRACE LITE Model: 75-7500-087 Post-Operative Knee Brace is indicated forJanuary 13, 2016
- FDANexGEN System REF 5980-37-01 Tibial Component Size Stemmed 3 Precoat, and REF 5986-37-01 Nonaugmentable TibialJanuary 13, 2016
- FDACorflex Contender Post-Op Knee Brace Lite Model: 75-7500-000 Post-Operative Knee Brace is indicated for rangeJanuary 13, 2016
- FDA3M Coban Self-Adherent Wrap Shipper carton label identified the product as 3M Coban LF Latex Free CatalogJanuary 13, 2016
- FDASmith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USEJanuary 13, 2016
- FDAGynecare Thermachoice III Uterine Balloon Therapy System Ethicon, Inc. The GYNECARE THERMACHOICE® III UterineJanuary 13, 2016
Questions about this recall
Was Remote Patient Monitoring System, MDP 2000 Series, Part Numbers: MDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected Examples of areas where the MDP200 is used include ICU, CCU, PACU, Emergency, Telemetry Step-down and other areas where patient's physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values which as heart rate and systolic, mean and diastolic pressure. This device acts as the Central Monitor Station, it permits from one (1) to sixteen (16) bedside monitor(s) to be connected simultaneously. recalled?
Yes. FDA published recall Z-0582-2016 on January 13, 2016. Reported hazard: Testing at customer site showed unit Remote Patient Monitoring System MDP2040-0100 in a continuous trap condition, not allowing system to reset and reboot. Two customers recently complained of the system freezing and it could only be restarted if the user re-applied power.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit