Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

Published January 2, 2013 · Recall initiated: June 6, 2006

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Radiometer America Inc
Distributor
Distributed to KY, NC, and NY.
Lot
Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.
Risk classification
Class III
Area (FDA)
device
Country
United States