AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers
Published October 11, 2017 · Recall initiated: September 7, 2017
Summary
FDA records medical device Class II recall Z-3237-2017, published October 11, 2017 and initiated September 7, 2017. Recalling firm: Radiometer America Inc; Status: Terminated; Reported quantity: 1669 Boxes; Lot: S7745 Lot R0698. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92 5.85 kPa or 36.9 46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Radiometer America Inc
- Distributor
- Worldwide Distribution - US (nationwide) Internationally to Canada, Netherlands, China, Czech Republic, Denmark, Germany, Spain France, Hungary, India, Denmark, Tokyo, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, and Finland.
- Lot
- S7745 Lot R0698
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAABL800 analyzer, model numbers 393-800 and 393-801.January 2, 2019
- FDAImplant Direct Swish Duo-Grip Healing Collar, Length: 3mm, Platform: 4.8 mm, Rx only, Sterile Dental ImplantOctober 11, 2017
- FDABOMImed Disposable Fiber Optic Medium Laryngoscope Handle. Each handle is packaged in an individual peelOctober 11, 2017
- FDAABL90 Flex Model #: 393-090 Portable, automated analyzer that measures pH, blood gases, electrolytes, glucose,March 22, 2017
- FDATCM CombiM Modules 903-111May 11, 2016
- FDAHemoCue® Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with theDecember 31, 2014
Questions about this recall
Was AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers recalled?
Yes. FDA published recall Z-3237-2017 on October 11, 2017. Reported hazard: Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92 5.85 kPa or 36.9 46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail