Access Immunoassay Systems Total T4 Calibrators, Part Number: 33805. The Access Total T4 Calibrators are intended to calibrate the Access Total T4 assay for the quantitative determination of total thyroxine (T4) levels in human serum and plasma using the Access Immunoassay Systems. Total T4 assay results are used for diagnosis and confirmation of thyroid disorders such as hyper- and hypothyroidism.

Published January 2, 2013 · Recall initiated: July 27, 2011

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A recall communication was initiated because Beckman has identified that Access Total T4 calibrator lots 019818, 021654, and 024072 do not meet their 12 month expiration date claim.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
Worldwide Distribution-USA (nationwide) and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Egypt, El Salvador, Ethiopia, France, Germany, Greece, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Lebanon, Libyan Arab Jamahiriya, Liechtenstein, Macao, Malaysia, Mexico, Netherlands, Nigeria, Pakistan, Panama, Philippines, Puerto Rico, Romania, Russian Federation, Serbia, Singapore, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia.
Lot
019818, 021654, 024072
Risk classification
Class II
Area (FDA)
device
Country
United States