GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B79632DA: EU: RT Refresh Package, Non-EU: Discovery RT, Optima CT580, or Discovery CT590 RT; X-ray/computed tomography system
Published December 11, 2024 · Recall initiated: November 5, 2024
Summary
FDA records medical device Class II recall Z-0613-2025, published December 11, 2024 and initiated November 5, 2024. Recalling firm: GE Healthcare (China) Co., Ltd.; Status: Ongoing; Reported quantity: 5 units; Lot: UDI/DI 00195278466488, Serial/Sales Order Numbers: 5469298, 5471981, 5461542, 5469371 4616154. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare (China) Co., Ltd.
- Distributor
- Worldwide distribution.
- Lot
- UDI/DI 00195278466488, Serial/Sales Order Numbers: 5469298, 5471981, 5461542, 5469371 4616154
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- China
Related recalls
- FDABrivo MR355, NMRI systemMarch 5, 2025
- FDAGE Healthcare Revolution Eagle, Model Number 5849504-3, X-ray/computed tomography systemDecember 11, 2024
- FDAGE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:December 11, 2024
- FDAGE Healthcare Optima 680 Expert, Model number 6670000; X-ray/computed tomography systemDecember 11, 2024
- FDAGE Healthcare Revolution Maxima Select, Model Number, 6670000-55; X-ray/computed tomography systemDecember 11, 2024
- FDAGE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80212DA: EU: Operator Console (forDecember 11, 2024
Questions about this recall
Was GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B79632DA: EU: RT Refresh Package, Non-EU: Discovery RT, Optima CT580, or Discovery CT590 RT; X-ray/computed tomography system recalled?
Yes. FDA published recall Z-0613-2025 on December 11, 2024. Reported hazard: GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
What should I do if I have this product?
Voluntary: Firm initiated · Letter