arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.
Published January 27, 2016 · Recall initiated: December 18, 2015
Summary
FDA records medical device Class I recall Z-0631-2016, published January 27, 2016 and initiated December 18, 2015. Recalling firm: Arkray Factory USA, Inc.; Status: Terminated; Reported quantity: 11 boxes (25 foil packaged test strips per box); Lot: Lot Number PN5C26. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Arkray Factory USA, Inc.
- Distributor
- Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.
- Lot
- Lot Number PN5C26
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrainlab Cranial Navigation System: An Image Guided Surgery System / Stereotactic. Radiology Departments. TheJanuary 27, 2016
- FDAVNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequencyJanuary 27, 2016
- FDABromfenac Ophthalmic Solution 0.09%, a) 2.5 mL bottles (NDC 0378-7110-35) and b) 5 mL bottles (NDCJanuary 27, 2016
- FDAInVision-Plus and InVision-Plus Junior Needleless IV Connector, Disposable, Single Use Products are intendedJanuary 27, 2016
- FDABevacizumab 2.5 mg/0.1 mL in 3/10 mL 31 G 5/16" Syringe for intravitreal injection, Rx -- PineJanuary 27, 2016
- FDAInVision-Plus CS Needleless IV Connector, Catalog Number RYM-7001. Disposable, Single Use. Products areJanuary 27, 2016
Questions about this recall
Was arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine. recalled?
Yes. FDA published recall Z-0631-2016 on January 27, 2016. Reported hazard: One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.
What should I do if I have this product?
Voluntary: Firm initiated · Letter