Brainlab Cranial Navigation System: An Image Guided Surgery System / Stereotactic. Radiology Departments. The BrainLAB Cranial IGS System is intended to be an intra-operative image guided localization system to enable minimally invasive surgery.
Published January 27, 2016 · Recall initiated: April 22, 2013
Summary
FDA records medical device Class I recall Z-0598-2016, published January 27, 2016 and initiated April 22, 2013. Recalling firm: Brainlab AG; Status: Terminated; Reported quantity: 2528 units / 2119 consignees; Lot: 1) Model 22035 , VVC NAVIGATION SOFTWARE CRANIAL 5.0 , Serial Numbers: 22035-103911-004 , 22035-103953-001 , 22035-103977-001 , 22035-103982-002 , 22035-103987-001 , 22035-103988-001 , 22035-104566-001 ; 2) Model 22040 , VV NAVIGATION SOFTWARE CRANIAL 4.0 , Serial Numbers: 22040/18820 , 22040/21370 , 22040/26909 , 22040/36997 , 22040-103675-006 , 22040-103687-009 , 22040-103710-001 , 22040-103942-008 , 22040-103974-011 , 22040-104067-001 , 22040-104072-001 , 22040-104114-001 , 22040-104148-001 , 22040-104151-001 , 22040-104505-001 , 22040-104542-001 , 22040-104545-001 , 22040-104614-005 , 22040-123772-001 ; 3) Model 22040-02 , VV NAV SW CRANIAL 7.8.1 (ANY PLATFORM) , Serial Number: 000000002600509001 ; 4) Model 22040-11 , UPDATE VV NAV SW CRANIAL 7.9.1 FROM 7.9 , Serial Numbers: 000000007665811001 , 000000007801111004 , 000000007801111005 , 000000007801111006 , 000000007801111007 , 000000007801111008 , 000000007801111009 , 000000007808711004 , 000000007808711007 , 000000007808711008 , 000000007837611010 , 000000007847411007 , 000000007865211004 , 000000007865211005 , 000000007942211004 , 000000007942211008 , 000000007942211010 , 000000007950411001 , 000000007950411004 , 000000007950411007 , 000000007950411010 , 000000007956611005 , 000000007956611010 , 000000007956611017 , 000000007965011001 , 000000007965011006 , 000000007965011007 , 000000007980111001 , 000000007980111002 , 000000007980111003 , 000000007980111004 , 000000007980111005 , 000000007980111006 , 000000007980111007 , 000000008000111003 , 000000008000111006 , 000000008000111007 , 000000008012411001 , 000000008012411003 , 000000008012411008 , 000000008026011002 , 000000008026011003 , 000000008036811002 , 000000008036811006 , 000000008036811007 , 000000008036811008 , 000000008036911002 , 000000008036911004 , 000000008036911005 , 000000008046111002 , 000000008046111005 , 000000008073511003 , 000000008073511007 , 000000008086811001 , 000000008106711001 , 000000008114111004 , 000000008129711004 , 000000008129711005 , 000000008129711006 , 000000008133311001 , 000000008133311002 , 000000008143111002 , 000000008143111003 , 000000008143111004 , 000000008157411005 , 000000008157411007 , 000000008167811005 , 000000008167811008 , 000000008167811012 , 000000008167811014 , 000000008167811015 , 000000008167811016 , 000000008180811001 , 000000008180811008 , 000000008180811009 , 000000008180811010 , 000000008180811011 , 000000008181011002 , 000000008208411002 , 000000008230111001 , 000000008230111002 , 000000008230111009 , 000000008230111013 , 000000008230111015 , 000000008230111022 , 000000008230111024 , 000000008230111025 , 000000008230211001 , 000000008230211002 , 000000008230211008 , 000000008230211017 , 000000008230211020 , 000000008230211025 , 000000008230211030 , 000000008230311006 , 000000008230311014 , 000000008230311015 , 000000008230311018 , 000000008230411003 , 000000008230411005 , 000000008230411009 , 000000008230411010 , 000000008230411012 , 000000008230411016 , 000000008230411021 , 000000008230411024 , 000000008230411029 , 000000008230411030 , 000000008230411031 , 000000008230411032 , 000000008242111002 , 000000008242111003 , 000000008242111005 , 000000008242611001 , 000000008242611002 , 000000009168012001 ; 5) Model 22040B , VV NAVIGATION SOFTWARE CRANIAL7.5.0 , Serial Numbers: 05-48916-22040B , 05-51803-22040B , 05-52149-22040B , 22040B/46911 ; 6) Model 22040C , VV NAVIGATION SOFTWARE CRANIAL 7.6.0 , Serial Numbers: 05-63659-22040C , 05-63669-22040C , 05-66754-22040C , 06-79914-22040C ; 7) Model 22040D , VV NAVIGATION SW CRANIAL 7.7 , Serial Numbers: 06-84302-22040D , 06-85635-22040D , 06-88801-22040D ; 8) Model 22040E , VV NAVIGATION SW CRANIAL 7.8 , Serial Numbers: 06-101594-22040E , 06-103819-22040E , 06-108132-22040E , 07-102189-22040E , 07-106354-22040E , 07-108464-22040E , 07-109071-. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Software Error: The effect of setup on overall navigation accuracy could potentially intensify small inaccuracies arising from individual steps of a complex navigation procedure that may cause an inaccurate display of instruments by the navigation system compared to the actual patient anatomy.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Brainlab AG
- Distributor
- Worldwide Distribution: US (Nationwide) including Puerto Rico; and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Latvia, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, and Vietnam.
- Lot
- 1) Model 22035 , VVC NAVIGATION SOFTWARE CRANIAL 5.0 , Serial Numbers: 22035-103911-004 , 22035-103953-001 , 22035-103977-001 , 22035-103982-002 , 22035-103987-001 , 22035-103988-001 , 22035-104566-001 ; 2) Model 22040 , VV NAVIGATION SOFTWARE CRANIAL 4.0 , Serial Numbers: 22040/18820 , 22040/21370 , 22040/26909 , 22040/36997 , 22040-103675-006 , 22040-103687-009 , 22040-103710-001 , 22040-103942-008 , 22040-103974-011 , 22040-104067-001 , 22040-104072-001 , 22040-104114-001 , 22040-104148-001 , 22040-104151-001 , 22040-104505-001 , 22040-104542-001 , 22040-104545-001 , 22040-104614-005 , 22040-123772-001 ; 3) Model 22040-02 , VV NAV SW CRANIAL 7.8.1 (ANY PLATFORM) , Serial Number: 000000002600509001 ; 4) Model 22040-11 , UPDATE VV NAV SW CRANIAL 7.9.1 FROM 7.9 , Serial Numbers: 000000007665811001 , 000000007801111004 , 000000007801111005 , 000000007801111006 , 000000007801111007 , 000000007801111008 , 000000007801111009 , 000000007808711004 , 000000007808711007 , 000000007808711008 , 000000007837611010 , 000000007847411007 , 000000007865211004 , 000000007865211005 , 000000007942211004 , 000000007942211008 , 000000007942211010 , 000000007950411001 , 000000007950411004 , 000000007950411007 , 000000007950411010 , 000000007956611005 , 000000007956611010 , 000000007956611017 , 000000007965011001 , 000000007965011006 , 000000007965011007 , 000000007980111001 , 000000007980111002 , 000000007980111003 , 000000007980111004 , 000000007980111005 , 000000007980111006 , 000000007980111007 , 000000008000111003 , 000000008000111006 , 000000008000111007 , 000000008012411001 , 000000008012411003 , 000000008012411008 , 000000008026011002 , 000000008026011003 , 000000008036811002 , 000000008036811006 , 000000008036811007 , 000000008036811008 , 000000008036911002 , 000000008036911004 , 000000008036911005 , 000000008046111002 , 000000008046111005 , 000000008073511003 , 000000008073511007 , 000000008086811001 , 000000008106711001 , 000000008114111004 , 000000008129711004 , 000000008129711005 , 000000008129711006 , 000000008133311001 , 000000008133311002 , 000000008143111002 , 000000008143111003 , 000000008143111004 , 000000008157411005 , 000000008157411007 , 000000008167811005 , 000000008167811008 , 000000008167811012 , 000000008167811014 , 000000008167811015 , 000000008167811016 , 000000008180811001 , 000000008180811008 , 000000008180811009 , 000000008180811010 , 000000008180811011 , 000000008181011002 , 000000008208411002 , 000000008230111001 , 000000008230111002 , 000000008230111009 , 000000008230111013 , 000000008230111015 , 000000008230111022 , 000000008230111024 , 000000008230111025 , 000000008230211001 , 000000008230211002 , 000000008230211008 , 000000008230211017 , 000000008230211020 , 000000008230211025 , 000000008230211030 , 000000008230311006 , 000000008230311014 , 000000008230311015 , 000000008230311018 , 000000008230411003 , 000000008230411005 , 000000008230411009 , 000000008230411010 , 000000008230411012 , 000000008230411016 , 000000008230411021 , 000000008230411024 , 000000008230411029 , 000000008230411030 , 000000008230411031 , 000000008230411032 , 000000008242111002 , 000000008242111003 , 000000008242111005 , 000000008242611001 , 000000008242611002 , 000000009168012001 ; 5) Model 22040B , VV NAVIGATION SOFTWARE CRANIAL7.5.0 , Serial Numbers: 05-48916-22040B , 05-51803-22040B , 05-52149-22040B , 22040B/46911 ; 6) Model 22040C , VV NAVIGATION SOFTWARE CRANIAL 7.6.0 , Serial Numbers: 05-63659-22040C , 05-63669-22040C , 05-66754-22040C , 06-79914-22040C ; 7) Model 22040D , VV NAVIGATION SW CRANIAL 7.7 , Serial Numbers: 06-84302-22040D , 06-85635-22040D , 06-88801-22040D ; 8) Model 22040E , VV NAVIGATION SW CRANIAL 7.8 , Serial Numbers: 06-101594-22040E , 06-103819-22040E , 06-108132-22040E , 07-102189-22040E , 07-106354-22040E , 07-108464-22040E , 07-109071-
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAUltrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x andOctober 14, 2020
- FDASpine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 )April 24, 2019
- FDAExacTrac versions 6.5 through 6.5 intended to be used to place patients at an accurately defined point withinMarch 21, 2018
- FDAAIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-rayFebruary 14, 2018
- FDABromfenac Ophthalmic Solution 0.09%, a) 2.5 mL bottles (NDC 0378-7110-35) and b) 5 mL bottles (NDCJanuary 27, 2016
- FDAarkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium,January 27, 2016
Questions about this recall
Was Brainlab Cranial Navigation System: An Image Guided Surgery System / Stereotactic. Radiology Departments. The BrainLAB Cranial IGS System is intended to be an intra-operative image guided localization system to enable minimally invasive surgery. recalled?
Yes. FDA published recall Z-0598-2016 on January 27, 2016. Reported hazard: Software Error: The effect of setup on overall navigation accuracy could potentially intensify small inaccuracies arising from individual steps of a complex navigation procedure that may cause an inaccurate display of instruments by the navigation system compared to the actual patient anatomy.
What should I do if I have this product?
Voluntary: Firm initiated · Letter