Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001
Published February 16, 2022 · Recall initiated: December 22, 2021
Summary
FDA records medical device Class II recall Z-0634-2022, published February 16, 2022 and initiated December 22, 2021. Recalling firm: Synthes (USA) Products LLC; Status: Ongoing; Reported quantity: 600 units; Lot: Lot/Serial No.: 20013101, 20052701, 20228901, 20229301, 20253901, 20558602, 20698102, 20777201 21869701, 21983904, 21995301 GTIN: 10886982297372. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes (USA) Products LLC
- Distributor
- US Nationwide distribution.
- Lot
- Lot/Serial No.: 20013101, 20052701, 20228901, 20229301, 20253901, 20558602, 20698102, 20777201 21869701, 21983904, 21995301 GTIN: 10886982297372
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASTARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screwJune 5, 2024
- FDA13 mm Titanium Hindfoot Arthrodesis Cannulated Nails EX, sterile -Intended to facilitate tibiotalocalcanealMarch 13, 2024
- FDARAPIDSORB IPS Battery Pack, Sterile-Intended for use in non-load bearing fracture repair and reconstructiveMarch 13, 2024
- FDAOpal, Intervertebral Cage, 10x32mm, h 11mm, Revolvable-Indicated for degenerative spine disease Part Number:March 13, 2024
- FDAATTUNE REVISION POSTERIOR FEMORAL AUGMENT SZ 9 12MM-intended as a Component in total knee arthroplasty (TKA)February 16, 2022
- FDALYSO Disk(TM) labeled as containing Actinomyces odontolyticus. Each LYSO Disk" unit contains six lyophilizedFebruary 16, 2022
Questions about this recall
Was Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001 recalled?
Yes. FDA published recall Z-0634-2022 on February 16, 2022. Reported hazard: Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.
What should I do if I have this product?
Voluntary: Firm initiated · Letter