Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110X11B EE2100X15B EE2110X15B ES2100X15B ES2110X15B EU2100X15B EU2100X19 EU2110X15B FR2100X14B FR2110X14B FX2100X15B GB2110X15B IA2100X15B IA2110X15B IN2100X15B IN2100X19 IN2110X15B IT2100X21B IT2110X21B JP2100X16B JP2110X16B KR2110X15B LA2100X15B LA2110X15B ND2110X15B RDE2100X13B RDE2110X13B RDS2110X11B RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B
Published January 10, 2024 · Recall initiated: November 27, 2023
Summary
FDA records medical device Class II recall Z-0638-2024, published January 10, 2024 and initiated November 27, 2023. Recalling firm: Philips Respironics, Inc.; Status: Ongoing; Reported quantity: 100,126 (54,691 US; 45,435 OUS) in total; Lot: Material UDI-DI/GTIN BL2110X15B 606959055520 BR2110X18B 606959055476 CA2110X12B 606959052307 DE2100X13B 606959054103 DE2110X13B 606959054110 DS2100X11B 606959051997 DS2110X11B 606959051942 EE2100X15B 606959055698 EE2110X15B 606959055544 ES2100X15B 606959054097 ES2110X15B 606959054080 EU2100X15B 606959055674 EU2100X19 606959055681 EU2110X15B 606959054981 FR2100X14B 606959055643 FR2110X14B 606959055506 FX2100X15B 606959061019 GB2110X15B 606959054127 IA2100X15B 606959055636 IA2110X15B 606959055490 IN2100X15B 606959054059 IN2100X19 606959055575 IN2110X15B 606959051959 IT2100X21B 606959055438 IT2110X21B 606959055421 JP2100X16B 606959055582 JP2110X16B 606959051973 KR2110X15B 606959055483 LA2100X15B 606959055599 LA2110X15B 606959055452 ND2110X15B 606959055513 RDE2100X13B RDE2110X13B RDS2110X11B 606959060517 RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B 606959060500 All units of listed models are affected by this correction. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Respironics, Inc.
- Distributor
- Worldwide distribution: US (nationwide)-AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Including PR; and OUS (International) including countries of: AE, AR,AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, ;EC EE, EG, ES, FI, FR, GB, GE, GF, GR, GT, HK, HR, HU, IE, IL, IN, IT, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, YT, ZA and ZW.
- Lot
- Material UDI-DI/GTIN BL2110X15B 606959055520 BR2110X18B 606959055476 CA2110X12B 606959052307 DE2100X13B 606959054103 DE2110X13B 606959054110 DS2100X11B 606959051997 DS2110X11B 606959051942 EE2100X15B 606959055698 EE2110X15B 606959055544 ES2100X15B 606959054097 ES2110X15B 606959054080 EU2100X15B 606959055674 EU2100X19 606959055681 EU2110X15B 606959054981 FR2100X14B 606959055643 FR2110X14B 606959055506 FX2100X15B 606959061019 GB2110X15B 606959054127 IA2100X15B 606959055636 IA2110X15B 606959055490 IN2100X15B 606959054059 IN2100X19 606959055575 IN2110X15B 606959051959 IT2100X21B 606959055438 IT2110X21B 606959055421 JP2100X16B 606959055582 JP2110X16B 606959051973 KR2110X15B 606959055483 LA2100X15B 606959055599 LA2110X15B 606959055452 ND2110X15B 606959055513 RDE2100X13B RDE2110X13B RDS2110X11B 606959060517 RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B 606959060500 All units of listed models are affected by this correction.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.April 1, 2026
- FDAPhilips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.April 1, 2026
- FDABiPAP A40 Ventilator. Used to provide noninvasive ventilation support for the purpose of treating adult andJuly 9, 2025
- FDAGarbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric throughNovember 20, 2024
- FDAImpella Connect, software portal for online device management that allows clinicians and Impella support staffJanuary 10, 2024
- FDAVitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation SystemJanuary 10, 2024
Questions about this recall
Was Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110X11B EE2100X15B EE2110X15B ES2100X15B ES2110X15B EU2100X15B EU2100X19 EU2110X15B FR2100X14B FR2110X14B FX2100X15B GB2110X15B IA2100X15B IA2110X15B IN2100X15B IN2100X19 IN2110X15B IT2100X21B IT2110X21B JP2100X16B JP2110X16B KR2110X15B LA2100X15B LA2110X15B ND2110X15B RDE2100X13B RDE2110X13B RDS2110X11B RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B recalled?
Yes. FDA published recall Z-0638-2024 on January 10, 2024. Reported hazard: The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.
What should I do if I have this product?
Voluntary: Firm initiated · Letter