Philips HeartStart FRx 861304 Automated External Defibrillator Manufactured by Philips Medical Systems, Seattle, WA 98121 USA; and Laerdal branded FRx 861305 The FRx Defibrillator is intended to treat ventricular fibrillation, the most common cause of sudden cardiac arrest. Using voice prompts, light emitting diodes (LEDs) and buttons, the user is guided through the response. The FRx uses a SMART biphasic, impedance compensating exponential waveform to deliver a nominal 150 J to adults and nominal 50 J to infants/children.
Published January 16, 2013 · Recall initiated: July 26, 2012
Summary
FDA records medical device Class II recall Z-0642-2013, published January 16, 2013 and initiated July 26, 2012. Recalling firm: Philips Medical Systems North America Co. Phillips; Status: Terminated; Reported quantity: 201,598; Lot: 1)Philips brand FRx Model Number 861304 under 510(k) K050004, sold nationally and Internationally, Serial numbers range from BO4L-00001 through B12F-00286 and represents 181,226 devices; 2) Laerdal brand FRx Model Number 861305 unfrt 510(k) K050004, sold nationally and Internationally, serial numbers range from B04L-00013 through B12E-04135 and represents 20,372 devices. Please note the serial number range represents first and last serial number for each model. Serial numbers are not continguous. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips determined that it is important to clarify information in the Owner's Manual and keep customers informed about the maintenance of their FRx (models 861304 and 861305), OnSite (models M5066A and M5067A), or HeartStart Home (model M5068A) automated external defibrillators (AEDs) shipped between December 2005 and July 2012. Philips had determined there is a need to emphasize the importance o
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems North America Co. Phillips
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of Japan and Malaysia.
- Lot
- 1)Philips brand FRx Model Number 861304 under 510(k) K050004, sold nationally and Internationally, Serial numbers range from BO4L-00001 through B12F-00286 and represents 181,226 devices; 2) Laerdal brand FRx Model Number 861305 unfrt 510(k) K050004, sold nationally and Internationally, serial numbers range from B04L-00013 through B12E-04135 and represents 20,372 devices. Please note the serial number range represents first and last serial number for each model. Serial numbers are not continguous.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips and Laerdal brands of HeartStart HS1 Defibrillator Family Model number M5066A, M5067A, and M5068AJanuary 16, 2013
- FDASiemens Chek-Stix Combo Pak (Distributed OUS) Catalog Number: 1364E Chek-Stix® Positive and Negative ControlJanuary 16, 2013
- FDASiemens Chek-Stix Combo Pak (Distributed OUS only) Catalog Number: 1364N Chek-Stix® Positive and NegativeJanuary 16, 2013
- FDALaparoscope, general; plastic surgery Used for general, abdominal, gynecological and thoracic minimallyJanuary 16, 2013
- FDASiemens Chek-Stix Combo Pak (Distributed OUS only) Catalog Number: 1364M Chek-Stix® Positive and NegativeJanuary 16, 2013
- FDAABX PENTRA ML, Software Versions: V8.0.x and Below; V.9.0.1 and V9.0.2January 16, 2013
Questions about this recall
Was Philips HeartStart FRx 861304 Automated External Defibrillator Manufactured by Philips Medical Systems, Seattle, WA 98121 USA; and Laerdal branded FRx 861305 The FRx Defibrillator is intended to treat ventricular fibrillation, the most common cause of sudden cardiac arrest. Using voice prompts, light emitting diodes (LEDs) and buttons, the user is guided through the response. The FRx uses a SMART biphasic, impedance compensating exponential waveform to deliver a nominal 150 J to adults and nominal 50 J to infants/children. recalled?
Yes. FDA published recall Z-0642-2013 on January 16, 2013. Reported hazard: Philips determined that it is important to clarify information in the Owner's Manual and keep customers informed about the maintenance of their FRx (models 861304 and 861305), OnSite (models M5066A and M5067A), or HeartStart Home (model M5068A) automated external defibrillators (AEDs) shipped between December 2005 and July 2012. Philips had determined there is a need to emphasize the importance o
What should I do if I have this product?
Voluntary: Firm initiated · Letter