Siemens Chek-Stix Combo Pak (Distributed OUS) Catalog Number: 1364E Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes
Published January 16, 2013 · Recall initiated: November 21, 2012
Summary
FDA records medical device Class III recall Z-0641-2013, published January 16, 2013 and initiated November 21, 2012. Recalling firm: Siemens Healthcare Diagnostics Inc; Status: Terminated; Reported quantity: 4642; Lot: Carton/Bottle Lot Number: C0099091C P0099091C C0100101F P0100101F C9101022I P9101022I C0103042D P0103042D C0106062E P0106062E. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25®C (77®F).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics Inc
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
- Lot
- Carton/Bottle Lot Number: C0099091C P0099091C C0100101F P0100101F C9101022I P9101022I C0103042D P0103042D C0106062E P0106062E
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;May 6, 2026
- FDARAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.August 6, 2025
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515November 13, 2024
- FDASoftware version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPointJuly 3, 2024
- FDASiemens Chek-Stix Combo Pak (Distributed OUS only) Catalog Number: 1364N Chek-Stix® Positive and NegativeJanuary 16, 2013
- FDAPhilips HeartStart FRx 861304 Automated External Defibrillator Manufactured by Philips Medical Systems,January 16, 2013
Questions about this recall
Was Siemens Chek-Stix Combo Pak (Distributed OUS) Catalog Number: 1364E Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes recalled?
Yes. FDA published recall Z-0641-2013 on January 16, 2013. Reported hazard: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25®C (77®F).
What should I do if I have this product?
Voluntary: Firm initiated · Letter