Siemens Chek-Stix Combo Pak (Distributed OUS) Catalog Number: 1364E Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes

Published January 16, 2013 · Recall initiated: November 21, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Healthcare Diagnostics Inc
Distributor
Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
Lot
Carton/Bottle Lot Number: C0099091C P0099091C C0100101F P0100101F C9101022I P9101022I C0103042D P0103042D C0106062E P0106062E
Risk classification
Class III
Area (FDA)
device
Country
United States