Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device with BZO, Reditest On-Site Oral Fluid Drug Screen Device & XALEX One Step Rapid Drug Test. Type of Packaging: 25 individually pouched test devices & collectors in a labeled kit box. The OrALert Oral Fluid Drug Screen Device for AMP/mAMP/COC/OPI/THC/PCP/BZO is an immunoassay based on the principle of competitive binding. Drugs that may be present in the oral fluid specimen compete against their respective drug conjugates for binding sites on their specific antibody.
Published January 15, 2014 · Recall initiated: June 3, 2013
Summary
FDA records medical device Class II recall Z-0653-2014, published January 15, 2014 and initiated June 3, 2013. Recalling firm: Alere San Diego, Inc.; Status: Terminated; Reported quantity: 280,710; Lot: Lot/Unit Codes: DSD-765-011 (Alere PN 4012350099) DOA2100709, DOA2110198, DOA2120093, DOA3010031, DOA3010095, DOA3010096, DOA3010393, DOA3010566, DOA3020420. DSD-765-011 (Alere PN 4080350099) DOA2100672, DOA2120094, DOA2120095, DOA2120158, DOA3010013, DOA3010064, DOA3010097, DOA3010394, DOA3010565, DOA3020557. DSF-765-031 (Alere PN 4584150099) DOA2100438, DOA2100667, DOA2110073, DOA2110315, DOA2120096, DOA2120738, DOA2120739, DOA3010003, DOA3010004, DOA3010005, DOA3010199, DOA3010202, DOA3020182, DOA3020183. DSF-765-011 (Alere PN 4582290164) DOA2100250, DOA2110316 , DOA2110317 , DOA2120502, DOA2120503, DOA3010236. DSF-765-011 (Alere PN 4028220211) DOA2100614. Expiration date, or Expected shelf life: 10/31/14 to 2/28/15. Expected shelf life: 24 months. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If the sponge becomes dislodged during the specimen collection process, the donor may inadvertently swallow the sponge which could present a choking risk.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Alere San Diego, Inc.
- Distributor
- US Distribution: including states of: AL, CA, PA, and VA.
- Lot
- Lot/Unit Codes: DSD-765-011 (Alere PN 4012350099) DOA2100709, DOA2110198, DOA2120093, DOA3010031, DOA3010095, DOA3010096, DOA3010393, DOA3010566, DOA3020420. DSD-765-011 (Alere PN 4080350099) DOA2100672, DOA2120094, DOA2120095, DOA2120158, DOA3010013, DOA3010064, DOA3010097, DOA3010394, DOA3010565, DOA3020557. DSF-765-031 (Alere PN 4584150099) DOA2100438, DOA2100667, DOA2110073, DOA2110315, DOA2120096, DOA2120738, DOA2120739, DOA3010003, DOA3010004, DOA3010005, DOA3010199, DOA3010202, DOA3020182, DOA3020183. DSF-765-011 (Alere PN 4582290164) DOA2100250, DOA2110316 , DOA2110317 , DOA2120502, DOA2120503, DOA3010236. DSF-765-011 (Alere PN 4028220211) DOA2100614. Expiration date, or Expected shelf life: 10/31/14 to 2/28/15. Expected shelf life: 24 months.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFisher Scientific Sure-Vue Urine hCG Strip, Model/ Part: FHC-101/23900527; McKesson CONSULT hCG Urine TestOctober 7, 2026
- FDAFisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + UrineOctober 7, 2026
- FDAMcKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine,October 7, 2026
- FDAFisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman CoulterOctober 7, 2026
- FDAArthrocare SPORTS MEDICINE, Knot Pusher, REF 25-3011, LOT #1049402, CON 1, 2013-10, NON-STERILE, Rx onlyJanuary 15, 2014
- FDA***REF BLACKMAX-N***Swivel/Angle Motor Assembly***Rx Only***The Anspach Effort, Inc. Intended for cutting andJanuary 15, 2014
Questions about this recall
Was Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device with BZO, Reditest On-Site Oral Fluid Drug Screen Device & XALEX One Step Rapid Drug Test. Type of Packaging: 25 individually pouched test devices & collectors in a labeled kit box. The OrALert Oral Fluid Drug Screen Device for AMP/mAMP/COC/OPI/THC/PCP/BZO is an immunoassay based on the principle of competitive binding. Drugs that may be present in the oral fluid specimen compete against their respective drug conjugates for binding sites on their specific antibody. recalled?
Yes. FDA published recall Z-0653-2014 on January 15, 2014. Reported hazard: Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If the sponge becomes dislodged during the specimen collection process, the donor may inadvertently swallow the sponge which could present a choking risk.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit