Sarns high-flow aortic arch cannula, 6.5 mm (20 Fr) OD with 3/8in connector, 11in (28 cm) long, with X coating surface The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery
Published January 23, 2013 · Recall initiated: January 7, 2013
Summary
FDA records medical device Class II recall Z-0671-2013, published January 23, 2013 and initiated January 7, 2013. Recalling firm: Terumo Cardiovascular Systems Corporation; Status: Terminated; Reported quantity: 6830 total all products; Lot: Lot number: 0677299 and Catalog number: 12325X. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Terumo Cardiovascular Systems Corporation
- Distributor
- Nationwide and worldwide: AL , AR , AZ , CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , ME , MI, MO , NC , ND , NV , NY , OH , OK , PA , SD , TN , TX , UT , VA , WA , and WI. Bahrain, BRAZIL, CANADA, CHILE, Honduras, Singapore, Taiwan, Trinidad and Tobago, and UNITED ARAB EMIRATES (UAE).
- Lot
- Lot number: 0677299 and Catalog number: 12325X
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPMJuly 2, 2025
- FDACardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974April 10, 2024
- FDA(1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, SterileDecember 23, 2020
- FDATerumo Capiox NX19 Oxygenator With Reservoir (East Orientation) P/N: 3CX*NX19REAugust 7, 2019
- FDASarns high-flow aortic arch cannula, 6.5 mm (20 Fr) OD with 3/8in connector, 11in (28 cm) long TheJanuary 23, 2013
- FDABausch and Lomb 27G sterile cannula packed in the Bausch and Lomb Amvisc 1.2% Sodium Hyaluronate ( ModelJanuary 23, 2013
Questions about this recall
Was Sarns high-flow aortic arch cannula, 6.5 mm (20 Fr) OD with 3/8in connector, 11in (28 cm) long, with X coating surface The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery recalled?
Yes. FDA published recall Z-0671-2013 on January 23, 2013. Reported hazard: Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.
What should I do if I have this product?
Voluntary: Firm initiated · Letter