FDAFDA · Food and drugs·Recall number Z-0672-2013

Bausch and Lomb 27G sterile cannula packed in the Bausch and Lomb Amvisc 1.2% Sodium Hyaluronate ( Model 59051, 59081, 59051L, 59081L) and Amvisc Plus 1.6% Sodium Hyaluronate (Model 60081, 60051, 60051L, 60081L) ophthalmic viscosurgical device. The 27G disposable cannula is attached to the syringe standard luer fitting to deliver the Amvis or Amvis Plus during use as a surgical aid in opthalmic anterior or posterior segment procedures.

Published January 23, 2013 · Recall initiated: November 15, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some disposable cannulas (with 4-digit lot numbers) provided with the Amvisc or Amvisc Plus OVD are not properly engaging to the Luer-Lock on the Amvisc and Amvisc Plus sterile glass syringes. The cannulas may leak viscoelastic material or detach from the syringe during injection. In rare incidents, detachment has resulted in serious patient injury.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Bausch and Lomb, Incorporated
Distributor
Worldwide distribution: USA including DC and PR; and countries of: Canada and Mexico accounts not available at this time.
Lot
Lot Numbers: 23632, 23633, 23639, 23640, 23647, 23801, 23811, 23863, 23886, 23897, 23920, 23933, 23972, 24022, 24025, 24028, 24257, 24323, 24436, B100003C, B100003E, B100003G, B100015A, B100015D, B100023A, B100023B, B100023C, B100023D, B100023E, B100039C, B100039D, B100039F, B100046A, B100046B, B100046C, B100058A, B100069A, B100069E, B110027D, B110028C, B110048A, B110048D, B110056A, B120011D, B120016A, B120046C, C100039A, C100039B, C10003B, C100046A, C100046B, C100046C, D090095B, D090096A, D090096C, D090097C, D090097D, D090097E, D090097G, D090105B, D090105C, D090105D, D100007B, D100007C, D100008A, D100008B, D100009A, D100011A, D100011B, D100011C, D100016A, D100016B, D100016C, D100016D, D100022A, D100022B, D100022C, D100022D, D100026A, D100026B, D100026C, D100026D, D100026E, D100030A, D100030B, D100030C, D100035A, D100035B, D100035C, D100035D, D100043A, D100043B, D100043C, D100044A, D100044B, D100044C, D100044D, D100044E, D100045A, D100045B, D100045C, D100045D, D100059A, D100059B, D100059C, D100060A, D100060B, D100061A, D100061B, D100080A, D100081A, D100081AR, D100081AS, D110015A, D110015B, D110016A, D110045A, D110046A, D110047A, D110070A, D110071A, D110072A, D110090A, D110090B, D110091A, D110092A, D110108A, D110108B, D110108BR, D120015A, D120015B, D120027A, D120028A, D120029A, D120029B, D120038B, D120039A, D120056B, D120068A, D120068B, E090105A, E100016A, E100016B, E100035A, E100035B, E100035C, E100045A, E100045B, E100059A, E100059B, E100060A, E100060B, E100060C, E100061A, Amvisc 1.2% Sodium Hyaluronate ( Model 59051, 59081, 59051L, 59081L) Amvisc Plus 1.6% Sodium Hyaluronate (Model 60081, 60051, 60051L, 60081L)
Risk classification
Class I
Area (FDA)
device
Country
United States