BOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use. Bond Polymer Refine Detection is a biotin-free, polymeric horseradish peroxidase (HRP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. It is intended for staining sections of formalin-fixed, paraffin-embedded tissue on the BOND automated system.
Published December 18, 2019 · Recall initiated: November 18, 2019
Summary
FDA records medical device Class II recall Z-0673-2020, published December 18, 2019 and initiated November 18, 2019. Recalling firm: Leica Microsystems, Inc.; Status: Ongoing; Reported quantity: 558; Lot: Lot: 65549. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
BOND Polymer Refine Detection kits (DS9800) were held out of refrigerated storage longer than specified which may exhibit suboptimal staining.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Leica Microsystems, Inc.
- Distributor
- Nationwide and OUS to Canada.
- Lot
- Lot: 65549
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABOND Ready-To-Use Primary Antibody CDX2 (EP25), REF PA0375September 22, 2021
- FDAElecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virusDecember 18, 2019
- FDAMaquet Cardiopulmonary GmbH / Getinge Quadrox-iD Pediatric Diffusion Membrane Oxygenator with Bioline Coating,December 18, 2019
- FDALeica M530 OHX Microscope System Supplied with 100 - 120 Vac.August 15, 2018
- FDANussloch GmbH ASP6025 Tissue ProcessorJuly 12, 2017
- FDANovocastra Liquid Mouse Monoclonal Antibody Calretinin, NCL-L-CALRET-566. Intended for the qualitativeJune 1, 2016
Questions about this recall
Was BOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use. Bond Polymer Refine Detection is a biotin-free, polymeric horseradish peroxidase (HRP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. It is intended for staining sections of formalin-fixed, paraffin-embedded tissue on the BOND automated system. recalled?
Yes. FDA published recall Z-0673-2020 on December 18, 2019. Reported hazard: BOND Polymer Refine Detection kits (DS9800) were held out of refrigerated storage longer than specified which may exhibit suboptimal staining.
What should I do if I have this product?
Voluntary: Firm initiated · Letter