Back to search
FDAFDA · Food and drugs·Recall number Z-1704-2016

Novocastra Liquid Mouse Monoclonal Antibody Calretinin, NCL-L-CALRET-566. Intended for the qualitative identification by light microscopy of human calretinin molecule in paraffin sections.

Published June 1, 2016 · Recall initiated: March 22, 2016

Summary

FDA records medical device Class II recall Z-1704-2016, published June 1, 2016 and initiated March 22, 2016. Recalling firm: Leica Microsystems, Inc.; Status: Terminated; Reported quantity: 290 units; Lot: Lot numbers: 6025204, 6029060, 6029949, 6035923 and 6036060. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Novocastra Liquid Mouse Monoclonal Antibody Calretinin when used in combination with specific Lots of antibody diluent will give reduced staining.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Leica Microsystems, Inc.
Distributor
Worldwide Distribution -- U.S., to the following states: AK, CA, FL, GA, MN, MT, NJ, OH, TX and WA; and, the country of Canada.
Lot
Lot numbers: 6025204, 6029060, 6029949, 6035923 and 6036060.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Novocastra Liquid Mouse Monoclonal Antibody Calretinin, NCL-L-CALRET-566. Intended for the qualitative identification by light microscopy of human calretinin molecule in paraffin sections. recalled?

Yes. FDA published recall Z-1704-2016 on June 1, 2016. Reported hazard: Novocastra Liquid Mouse Monoclonal Antibody Calretinin when used in combination with specific Lots of antibody diluent will give reduced staining.

What should I do if I have this product?

Voluntary: Firm initiated · Letter