DX-D 600 DIRECT RADIOGRAPHY SYSTEM Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.
Published January 16, 2019 · Recall initiated: December 4, 2018
Summary
FDA records medical device Class II recall Z-0692-2019, published January 16, 2019 and initiated December 4, 2018. Recalling firm: Agfa-Gevaert, N.V.; Status: Terminated; Reported quantity: 10; Lot: Serial Numbers: A5430000015 A5430000020 A5430000030 A5430000042 A5430000052 A5430000061 A5430000066 A5430000067 A5430000079 A5430000081. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
After an upgrade of the software of the Overhead Tube Crane, there were isolated cases in which the Overhead Tube Crane movement does not stop when the movement button is released. Instead of stopping while the movement button is released, the Tube Head Crane moves to the intended position.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Agfa-Gevaert, N.V.
- Distributor
- US Nationwide Distribution in the states: IN, KS, MA, SC, and VA.
- Lot
- Serial Numbers: A5430000015 A5430000020 A5430000030 A5430000042 A5430000052 A5430000061 A5430000066 A5430000067 A5430000079 A5430000081
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Belgium
Related recalls
- FDAAmlodipine and Valsartan Tablets USP 10 mg/320 mg. 30-count bottles, Rx Only, Manufactured for: AurobindoJanuary 23, 2019
- FDAAdam Bros. Family Farms Cauliflower; Sold in cases with 9, 12, 16 or 20 count casesJanuary 16, 2019
- FDAddRCombi Plus FP, Model Number 8.2000.0070.0 (US Version), 8.2000.0060.0 (Europe Version) The ddRCombi Plus FPJanuary 16, 2019
- FDAAdam Bros. Family Farms Green Leaf Lettuce; 24 heads per caseJanuary 16, 2019
- FDAIntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when usedJanuary 16, 2019
- FDAAdam Bros. Family Farms Red Leaf Lettuce 24 heads per carton Adams Bros Family Farms and OvationJanuary 16, 2019
Questions about this recall
Was DX-D 600 DIRECT RADIOGRAPHY SYSTEM Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients. recalled?
Yes. FDA published recall Z-0692-2019 on January 16, 2019. Reported hazard: After an upgrade of the software of the Overhead Tube Crane, there were isolated cases in which the Overhead Tube Crane movement does not stop when the movement button is released. Instead of stopping while the movement button is released, the Tube Head Crane moves to the intended position.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail