Coated VICRYL RAPIDETM (Polyglactin 910) Suture
Published March 7, 2018 · Recall initiated: August 17, 2017
Summary
FDA records medical device Class II recall Z-0693-2018, published March 7, 2018 and initiated August 17, 2017. Recalling firm: Ethicon, Inc.; Status: Terminated; Reported quantity: 129,372 Eaches (Internationally); Lot: Coated VICRYL RAPIDETM (Polyglactin 910) Suture MPVR4260H Size 4-0 VICRYL RAPIDE Undyed 75cm PS-2 Needle 10705031149168 Coated VICRYL RAPIDETM (Polyglactin 910) Suture V2140H Size 4-0 VICRYL RAPIDE Undyed 70cm RB-1 Needle 10705031106048 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2217 Size 2-0 VICRYL RAPIDE Undyed 90cm V-34 Needle 10705031119925 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2253 Size 2-0 VICRYL RAPIDE Undyed 75cm FS-1 Needle 10705031119956 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2279 Size 4-0 VICRYL RAPIDE Undyed 75cm V-4 Needle 10705031086524 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2294 Size 4-0 VICRYL RAPIDE Undyed 75cm FS-3 Needle 10705031086548 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2297 Size 4-0 VICRYL RAPIDE Undyed 75cm FS-2 Needle 10705031119970 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR496 Size 4-0 VICRYL RAPIDE Undyed 18'' PS-2 Needle 10705031134690 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR945 Size 2-0 VICRYL RAPIDE Undyed 36'' CT-1 Needle 10705031133358 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI253H Size 2-0 VICRYL RAPIDE Undyed 75cm FS-1 Needle 10705031111172 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9941 Size 2-0 VICRYL RAPIDE Undyed 75cm FS Needle 10705031130999 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR426 Size 4-0 VICRYL RAPIDE Undyed 75cm PS-2 Needle 10705031134669 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR494 Size 4-0 VICRYL RAPIDE Undyed 45cm P-3 Needle 10705031134683 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR845 Size 4-0 VICRYL RAPIDE Undyed 45cm PC-3 Needle 10705031134744 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR917 Size 2-0 VICRYL RAPIDE Undyed 90cm V-34 Needle 10705031134768 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2286 Size 2-0 VICRYL RAPIDE Undyed 90cm FSL Needle 10705031080256 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2299 Size 2-0 VICRYL RAPIDE Undyed 75cm FS-2 Needle 10705031080294 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9918 Size 4-0 VICRYL RAPIDE Undyed 75cm PC-3 Needle 10705031151970 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9930 Size 4-0 VICRYL RAPIDE Undyed 75cm FS-2 Needle 10705031152007 Coated VICRYL RAPIDETM (Polyglactin 910) Suture FH1642H Size 4-0 VICRYL RAPIDE Undyed 45cm FS-2 Needle 10705031099968 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI214H Size 4-0 VICRYL RAPIDE Undyed 75cm RB-1 Plus Needle 10705031111127 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI312H Size 2-0 VICRYL RAPIDE Undyed 75cm SH-1 Plus Needle 10705031111264 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI417H Size 2-0 VICRY RAPIDE Undyed 75cm SH Plus Needle 10705031111332 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9922 Size 4-0 VICRYL RAPIDE Undyed 75cm PS-2 Needle 10705031130845 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9933 Size 2-0 VICRYL RAPIDE Undyed 75cm PS Needle 10705031130937. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
On August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture that were distributed to the field. During finished goods testing, a suture damage was identified on a lot of VICRYL RAPIDETM Suture. During the course of the subsequent investigation, the suture also failed knot tensile strength testing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ethicon, Inc.
- Distributor
- Nationally
- Lot
- Coated VICRYL RAPIDETM (Polyglactin 910) Suture MPVR4260H Size 4-0 VICRYL RAPIDE Undyed 75cm PS-2 Needle 10705031149168 Coated VICRYL RAPIDETM (Polyglactin 910) Suture V2140H Size 4-0 VICRYL RAPIDE Undyed 70cm RB-1 Needle 10705031106048 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2217 Size 2-0 VICRYL RAPIDE Undyed 90cm V-34 Needle 10705031119925 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2253 Size 2-0 VICRYL RAPIDE Undyed 75cm FS-1 Needle 10705031119956 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2279 Size 4-0 VICRYL RAPIDE Undyed 75cm V-4 Needle 10705031086524 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2294 Size 4-0 VICRYL RAPIDE Undyed 75cm FS-3 Needle 10705031086548 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2297 Size 4-0 VICRYL RAPIDE Undyed 75cm FS-2 Needle 10705031119970 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR496 Size 4-0 VICRYL RAPIDE Undyed 18'' PS-2 Needle 10705031134690 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR945 Size 2-0 VICRYL RAPIDE Undyed 36'' CT-1 Needle 10705031133358 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI253H Size 2-0 VICRYL RAPIDE Undyed 75cm FS-1 Needle 10705031111172 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9941 Size 2-0 VICRYL RAPIDE Undyed 75cm FS Needle 10705031130999 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR426 Size 4-0 VICRYL RAPIDE Undyed 75cm PS-2 Needle 10705031134669 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR494 Size 4-0 VICRYL RAPIDE Undyed 45cm P-3 Needle 10705031134683 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR845 Size 4-0 VICRYL RAPIDE Undyed 45cm PC-3 Needle 10705031134744 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR917 Size 2-0 VICRYL RAPIDE Undyed 90cm V-34 Needle 10705031134768 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2286 Size 2-0 VICRYL RAPIDE Undyed 90cm FSL Needle 10705031080256 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2299 Size 2-0 VICRYL RAPIDE Undyed 75cm FS-2 Needle 10705031080294 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9918 Size 4-0 VICRYL RAPIDE Undyed 75cm PC-3 Needle 10705031151970 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9930 Size 4-0 VICRYL RAPIDE Undyed 75cm FS-2 Needle 10705031152007 Coated VICRYL RAPIDETM (Polyglactin 910) Suture FH1642H Size 4-0 VICRYL RAPIDE Undyed 45cm FS-2 Needle 10705031099968 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI214H Size 4-0 VICRYL RAPIDE Undyed 75cm RB-1 Plus Needle 10705031111127 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI312H Size 2-0 VICRYL RAPIDE Undyed 75cm SH-1 Plus Needle 10705031111264 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI417H Size 2-0 VICRY RAPIDE Undyed 75cm SH Plus Needle 10705031111332 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9922 Size 4-0 VICRYL RAPIDE Undyed 75cm PS-2 Needle 10705031130845 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9933 Size 2-0 VICRYL RAPIDE Undyed 75cm PS Needle 10705031130937
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACoated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CT, Product Number J358HNovember 13, 2024
- FDAVICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation,June 26, 2024
- FDAPDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation,June 26, 2024
- FDAMONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligationJune 26, 2024
- FDABiofinity XR Toric contact lenses 8.7/14.5 - 6PK Biofinity XR Toric (comfilcon A) Soft (Hydrophilic) ContactMarch 7, 2018
- FDAFootswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage:March 7, 2018
Questions about this recall
Was Coated VICRYL RAPIDETM (Polyglactin 910) Suture recalled?
Yes. FDA published recall Z-0693-2018 on March 7, 2018. Reported hazard: On August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture that were distributed to the field. During finished goods testing, a suture damage was identified on a lot of VICRYL RAPIDETM Suture. During the course of the subsequent investigation, the suture also failed knot tensile strength testing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter