Revolution CT with Apex Edition CT System - Product Usage: intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.
Published December 30, 2020 · Recall initiated: November 23, 2020
Summary
FDA records medical device Class II recall Z-0695-2021, published December 30, 2020 and initiated November 23, 2020. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 3 units; Lot: Model Number: 5995000-5; Serial Number's: REVVX1700078CN (System ID: 713MMREVCT/UDI: 01008406821185521117070021REVVX1700078CN) , 00000440315CN0 (System ID: 220078CT04), and REVVX1800040CN (System ID: M2322265/UDI: 01008406821185521118030021REVVX1800040CN). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A potential for protocol(s) with unintended patient orientation or scan setting when using Intelligent Protocoling
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide distribution - US Nationwide distribution including in the states of FL and TX. The countries of Belgium, France, and Spain.
- Lot
- Model Number: 5995000-5; Serial Number's: REVVX1700078CN (System ID: 713MMREVCT/UDI: 01008406821185521117070021REVVX1700078CN) , 00000440315CN0 (System ID: 220078CT04), and REVVX1800040CN (System ID: M2322265/UDI: 01008406821185521118030021REVVX1800040CN).
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDARevolution CT ES - Product Usage: intended for head, whole body, cardiac and vascular X-ray ComputedDecember 30, 2020
- FDASpinal Rod Cutter, Item number 00392500200. The product is a nonsterile pin cutter - Product Usage: TheDecember 30, 2020
Questions about this recall
Was Revolution CT with Apex Edition CT System - Product Usage: intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. recalled?
Yes. FDA published recall Z-0695-2021 on December 30, 2020. Reported hazard: A potential for protocol(s) with unintended patient orientation or scan setting when using Intelligent Protocoling
What should I do if I have this product?
Voluntary: Firm initiated · Letter