Soltive SuperPulsed Laser System TFL Fiber Stripper, Model Numbers TFL-AFS150, TFL-AFS200, TFL-AFS365, TFL-AFS550, TFL-AFS940
Published January 24, 2024 · Recall initiated: December 4, 2023
Summary
FDA records medical device Class II recall Z-0716-2024, published January 24, 2024 and initiated December 4, 2023. Recalling firm: Olympus Corporation of the Americas; Status: Ongoing; Reported quantity: 3001 units; Lot: Lot Number: All units sold prior to 2023. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- US Nationwide. Global Distribution
- Lot
- Lot Number: All units sold prior to 2023
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.October 7, 2026
- FDAOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.October 7, 2026
- FDAOlympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunctionOctober 7, 2026
- FDAOlympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.September 30, 2026
- FDAProduct REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery UnitJanuary 24, 2024
- FDASoltive SuperPulsed Laser System TFL FIBER CLEAVER, Model Number TFL-AFCJanuary 24, 2024
Questions about this recall
Was Soltive SuperPulsed Laser System TFL Fiber Stripper, Model Numbers TFL-AFS150, TFL-AFS200, TFL-AFS365, TFL-AFS550, TFL-AFS940 recalled?
Yes. FDA published recall Z-0716-2024 on January 24, 2024. Reported hazard: The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.
What should I do if I have this product?
Voluntary: Firm initiated · Letter