FDAFDA · Food and drugs·Recall number Z-0720-2021

MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic cocci and listeria monocytogenes.

Published December 30, 2020 · Recall initiated: December 2, 2020

Summary

FDA records medical device Class II recall Z-0720-2021, published December 30, 2020 and initiated December 2, 2020. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 685 panel boxes (20 panels per box) = 13,700 panels; Lot: Catalog Number: B1017-216 UDI Code: (01)15099590657550(17)210610(11)200610(10)2021-06-10 Lot Number: 2021-06-10. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
U.S. Nationwide distribution including in the states of AL, AZ, CA, FL, IN, KY, MA, MI, MO, NC, NH, NY, OH, OK, PA, SC, SD, TN, TX, VA, and WI. The countries of Australia, Austria, Bahamas, Bangladesh, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura®ao,¿St®Eus, Czech®Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, India, Iran, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Uruguay, Vatican City, Vietnam.
Lot
Catalog Number: B1017-216 UDI Code: (01)15099590657550(17)210610(11)200610(10)2021-06-10 Lot Number: 2021-06-10
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic cocci and listeria monocytogenes. recalled?

Yes. FDA published recall Z-0720-2021 on December 30, 2020. Reported hazard: Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.

What should I do if I have this product?

Voluntary: Firm initiated · Letter