Curaplex Cricothyrotomy Field Kit with Instruments, Kwik Cric MVP, 6.0mm, Item Number 351631i-A. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy.
Published January 24, 2024 · Recall initiated: November 6, 2023
Summary
FDA records medical device Class II recall Z-0721-2024, published January 24, 2024 and initiated November 6, 2023. Recalling firm: Sarnova HC, Llc; Status: Ongoing; Reported quantity: 15 kits; Lot: Lot Numbers: ASM0037685 ASM0037887. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sarnova HC, Llc
- Distributor
- US Nationwide
- Lot
- Lot Numbers: ASM0037685 ASM0037887
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACuraplex Capnography EtCO2/O2 Oral/Nasal Cannula Multipack (10 pk), Item Number 670272-KITOctober 21, 2026
- FDACuraplex Nasal Intubation Kit w/ 7mm Endotracheal Tube-To perform a nasal intubation to establish an airway inMarch 27, 2024
- FDACuraplex Nasal Intubation Kit with 6mm Endotrol Tube-To perform a nasal intubation to establish an airway inMarch 27, 2024
- FDACuraplex Field Cric Kit with ET Tube, Item Number 36611. The kit provides healthcare professionals with anJanuary 24, 2024
- FDAMINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429January 24, 2024
- FDAIncisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cable part numberJanuary 24, 2024
Questions about this recall
Was Curaplex Cricothyrotomy Field Kit with Instruments, Kwik Cric MVP, 6.0mm, Item Number 351631i-A. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy. recalled?
Yes. FDA published recall Z-0721-2024 on January 24, 2024. Reported hazard: It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.
What should I do if I have this product?
Voluntary: Firm initiated · Letter