MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
Published January 24, 2024 · Recall initiated: December 7, 2023
Summary
FDA records medical device Class II recall Z-0722-2024, published January 24, 2024 and initiated December 7, 2023. Recalling firm: Cook Medical Llc; Status: Ongoing; Reported quantity: 14 US; 101 OUS; Lot: UDI-DI 00827002444294 Lots AM109365 AM109363 AM109356 AM111934 AM111932 AM111933 AM112041 AM112042 AM112039 AM112040 AM112043 AM112071 AM112072 AM112316. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Medical Llc
- Distributor
- Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
- Lot
- UDI-DI 00827002444294 Lots AM109365 AM109363 AM109356 AM111934 AM111932 AM111933 AM112041 AM112042 AM112039 AM112040 AM112043 AM112071 AM112072 AM112316
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACuraplex Cricothyrotomy Field Kit with Instruments, Kwik Cric MVP, 6.0mm, Item Number 351631i-A. The kitJanuary 24, 2024
- FDAVasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3000-W .January 24, 2024
- FDACuraplex Field Cric Kit with ET Tube, Item Number 36611. The kit provides healthcare professionals with anJanuary 24, 2024
- FDAVasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3500.January 24, 2024
- FDAIncisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cable part numberJanuary 24, 2024
- FDASpectral CT 7500, Software Version 5.0.0.X. Computed Tomography X-Ray System.January 24, 2024
Questions about this recall
Was MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429 recalled?
Yes. FDA published recall Z-0722-2024 on January 24, 2024. Reported hazard: The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
What should I do if I have this product?
Voluntary: Firm initiated · Letter