GE Optima XR220amx and Optima XR200amx with Digital Upgrade Mobile Radiographic Systems. Model numbers 5555000-5, 5555000-6, and 5421698.
Published January 22, 2014 · Recall initiated: November 7, 2013
Summary
FDA records medical device Class II recall Z-0725-2014, published January 22, 2014 and initiated November 7, 2013. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 1027; Lot: Serial Numbers: 00000002796EN1, 00000002797EN9, 00000002798EN7, 00000002799EN5, 00000002800EN1, 00000002801EN9, 00000102726WK9, 00000102789WK3, 00000102937WK7, 00000130778WK3, 00000UA45526-8, 00001027058WK5, 00001027061WK9, 00001027146WK8, 00001027149WK2, 00001027150WK0, 00001027151WK8, 00001027153WK4, 00001027154WK2, 00001027167WK4, 00001027168WK2, 00001027169WK0, 00001027175WK7, 00001027193WK0, 00001027194WK8, 00001027195WK5, 00001027199WK7, 00001027200WK3, 00001027203WK7, 00001027204WK5, 00001027207WK8, 00001027214WK4, 00001027256WK5, 00001027259WK9, 00001027260WK7, 00001027262WK3, 00001027264WK9, 00001027265WK6, 00001027267WK2, 00001027268WK0, 00001027269WK8, 00001027276WK3, 00001027277WK1, 00001027278WK9, 00001027280WK5, 00001027293WK8, 00001027294WK6, 00001027295WK3, 00001027297WK9, 00001027298WK7, 00001027299WK5, 00001027300WK1, 00001027301WK9, 00001027302WK7, 00001027303WK5, 00001027314WK2, 00001027318WK3, 00001027322WK5, 00001027323WK3, 00001027324WK1, 00001027325WK8, 00001027326WK6, 00001027344WK9, 00001027346WK4, 00001027347WK2, 00001027349WK8, 00001027350WK6, 00001027351WK4, 00001027352WK2, 00001027353WK0, 00001027354WK8, 00001027356WK3, 00001027359WK7, 00001027362WK1, 00001027363WK9, 00001027364WK7, 00001027369WK6, 00001027371WK2, 00001027372WK0, 00001027373WK8, 00001027374WK6, 00001027376WK1, 00001027377WK9, 00001027380WK3, 00001027382WK9, 00001027383WK7, 00001027384WK5, 00001027386WK0, 00001027387WK8, 00001027388WK6, 00001027389WK4, 00001027390WK2, 00001027406WK6, 00001027407WK4, 00001027416WK5, 00001027417WK3, 00001027418WK1, 00001027421WK5, 00001027422WK3, 00001027423WK1, 00001027438WK9, 00001027445WK4, 00001027450WK4, 00001027451WK2, 00001027452WK0, 00001027464WK5, 00001027465WK2, 00001027466WK0, 00001027487WK6, 00001027488WK4, 00001027489WK2, 00001027490WK0, 00001027506WK3, 00001027507WK1, 00001027508WK9, 00001027539WK4, 00001027540WK2, 00001027541WK0, 00001027542WK8, 00001027543WK6, 00001027544WK4, 00001027551WK9, 00001027552WK7, 00001027553WK5, 00001027582WK4, 00001027583WK2, 00001027699WK6, 00001027700WK2, 00001027736WK6, 00001027737WK4, 00001027749WK9, 00001027762WK2, 00001027764WK8, 00001027765WK5, 00001027766WK3, 00001027768WK9, 00001027769WK7, 00001027772WK1, 00001027773WK9, 00001027774WK7, 00001027775WK4, 00001027778WK8, 00001027779WK6, 00001027781WK2, 00001027814WK1, 00001027819WK0, 00001027820WK8, 00001027821WK6, 00001027823WK2, 00001027824WK0, 00001027825WK7, 00001027826WK5, 00001027827WK3, 00001027828WK1, 00001027878WK6, 00001027880WK2, 00001027887WK7, 00001027888WK5, 00001027889WK3, 00001027890WK1, 00001027891WK9, 00001027892WK7, 00001027893WK5, 00001027894WK3, 00001027895WK0, 00001027896WK8, 00001027915WK6, 00001027932WK1, 00001027934WK7, 00001027935WK4, 00001027939WK6, 00001027992WK5, 00001027993WK3, 00001027994WK1, 00001027995WK8, 00001027996WK6, 00001028003WK0, 00001028021WK2, 00001028022WK0, 00001028023WK8, 00001028025WK3, 00001028035WK2, 00001028036WK0, 00001028042WK8, 00001028043WK6, 00001028044WK4, 00001028050WK1, 00001028051WK9, 00001028052WK7, 00001028053WK5, 00001028064WK2, 00001028065WK9, 00001028066WK7, 00001028077WK4, 00001028078WK2, 00001028079WK0, 00001028080WK8, 00001028091WK5, 00001028092WK3, 00001028093WK1, 00001028094WK9, 00001028119WK4, 00001028120WK2, 00001028205WK1, 00001028213WK5, 00001028214WK3, 00001028215WK0, 00001028216WK8, 00001028217WK6, 00001028230WK9, 00001028231WK7, 00001028232WK5, 00001028267WK1, 00001028269WK7, 00001028275WK4, 00001028276WK2, 00001028277WK0, 00001028279WK6, 00001028280WK4, 00001028281WK2, 00001028283WK8, 00001028284WK6, 00001028285WK3, 00001028286WK1, 00001028287WK9, 00001028288WK7, 00001028289WK5, 00001028290WK3, 00001028291WK1, 00001028292WK9, 00001028294WK5, 00001028331WK5, 00001028333WK1, 00001028334WK9, 00001028337WK2, 00001028338WK0, 00001028339WK8, 00001028340WK6, 00001028341WK4, 00001028342WK2, 00001028343WK0, 00001028344WK8, 00001028345WK5, 00001028346WK3, 00001028347WK1, 00001028351WK3, 00001028352WK1, 00001028353WK9, 00001028354WK7,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There exists a software issue associated with patient selection from the worklist on the Optima XR200 with Digital Upgrade, and Optima XR220amx Mobile X-ray Systems. Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the mobile x-ray system are not properly displayed. If the user continues to attempt to select one of the exams w
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution-USA (nationwide) including Washington D.C., and Guam and the countries of ALBANIA, AUSTRALIA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CHILE, CHINA , COLOMBIA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, IRAQ, IRELAND, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MEXICO, NEW ZEALAND, NIGERIA, NORWAY, POLAND, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, UNITED ARAB EMIRATES, and UNITED KINGDOM.
- Lot
- Serial Numbers: 00000002796EN1, 00000002797EN9, 00000002798EN7, 00000002799EN5, 00000002800EN1, 00000002801EN9, 00000102726WK9, 00000102789WK3, 00000102937WK7, 00000130778WK3, 00000UA45526-8, 00001027058WK5, 00001027061WK9, 00001027146WK8, 00001027149WK2, 00001027150WK0, 00001027151WK8, 00001027153WK4, 00001027154WK2, 00001027167WK4, 00001027168WK2, 00001027169WK0, 00001027175WK7, 00001027193WK0, 00001027194WK8, 00001027195WK5, 00001027199WK7, 00001027200WK3, 00001027203WK7, 00001027204WK5, 00001027207WK8, 00001027214WK4, 00001027256WK5, 00001027259WK9, 00001027260WK7, 00001027262WK3, 00001027264WK9, 00001027265WK6, 00001027267WK2, 00001027268WK0, 00001027269WK8, 00001027276WK3, 00001027277WK1, 00001027278WK9, 00001027280WK5, 00001027293WK8, 00001027294WK6, 00001027295WK3, 00001027297WK9, 00001027298WK7, 00001027299WK5, 00001027300WK1, 00001027301WK9, 00001027302WK7, 00001027303WK5, 00001027314WK2, 00001027318WK3, 00001027322WK5, 00001027323WK3, 00001027324WK1, 00001027325WK8, 00001027326WK6, 00001027344WK9, 00001027346WK4, 00001027347WK2, 00001027349WK8, 00001027350WK6, 00001027351WK4, 00001027352WK2, 00001027353WK0, 00001027354WK8, 00001027356WK3, 00001027359WK7, 00001027362WK1, 00001027363WK9, 00001027364WK7, 00001027369WK6, 00001027371WK2, 00001027372WK0, 00001027373WK8, 00001027374WK6, 00001027376WK1, 00001027377WK9, 00001027380WK3, 00001027382WK9, 00001027383WK7, 00001027384WK5, 00001027386WK0, 00001027387WK8, 00001027388WK6, 00001027389WK4, 00001027390WK2, 00001027406WK6, 00001027407WK4, 00001027416WK5, 00001027417WK3, 00001027418WK1, 00001027421WK5, 00001027422WK3, 00001027423WK1, 00001027438WK9, 00001027445WK4, 00001027450WK4, 00001027451WK2, 00001027452WK0, 00001027464WK5, 00001027465WK2, 00001027466WK0, 00001027487WK6, 00001027488WK4, 00001027489WK2, 00001027490WK0, 00001027506WK3, 00001027507WK1, 00001027508WK9, 00001027539WK4, 00001027540WK2, 00001027541WK0, 00001027542WK8, 00001027543WK6, 00001027544WK4, 00001027551WK9, 00001027552WK7, 00001027553WK5, 00001027582WK4, 00001027583WK2, 00001027699WK6, 00001027700WK2, 00001027736WK6, 00001027737WK4, 00001027749WK9, 00001027762WK2, 00001027764WK8, 00001027765WK5, 00001027766WK3, 00001027768WK9, 00001027769WK7, 00001027772WK1, 00001027773WK9, 00001027774WK7, 00001027775WK4, 00001027778WK8, 00001027779WK6, 00001027781WK2, 00001027814WK1, 00001027819WK0, 00001027820WK8, 00001027821WK6, 00001027823WK2, 00001027824WK0, 00001027825WK7, 00001027826WK5, 00001027827WK3, 00001027828WK1, 00001027878WK6, 00001027880WK2, 00001027887WK7, 00001027888WK5, 00001027889WK3, 00001027890WK1, 00001027891WK9, 00001027892WK7, 00001027893WK5, 00001027894WK3, 00001027895WK0, 00001027896WK8, 00001027915WK6, 00001027932WK1, 00001027934WK7, 00001027935WK4, 00001027939WK6, 00001027992WK5, 00001027993WK3, 00001027994WK1, 00001027995WK8, 00001027996WK6, 00001028003WK0, 00001028021WK2, 00001028022WK0, 00001028023WK8, 00001028025WK3, 00001028035WK2, 00001028036WK0, 00001028042WK8, 00001028043WK6, 00001028044WK4, 00001028050WK1, 00001028051WK9, 00001028052WK7, 00001028053WK5, 00001028064WK2, 00001028065WK9, 00001028066WK7, 00001028077WK4, 00001028078WK2, 00001028079WK0, 00001028080WK8, 00001028091WK5, 00001028092WK3, 00001028093WK1, 00001028094WK9, 00001028119WK4, 00001028120WK2, 00001028205WK1, 00001028213WK5, 00001028214WK3, 00001028215WK0, 00001028216WK8, 00001028217WK6, 00001028230WK9, 00001028231WK7, 00001028232WK5, 00001028267WK1, 00001028269WK7, 00001028275WK4, 00001028276WK2, 00001028277WK0, 00001028279WK6, 00001028280WK4, 00001028281WK2, 00001028283WK8, 00001028284WK6, 00001028285WK3, 00001028286WK1, 00001028287WK9, 00001028288WK7, 00001028289WK5, 00001028290WK3, 00001028291WK1, 00001028292WK9, 00001028294WK5, 00001028331WK5, 00001028333WK1, 00001028334WK9, 00001028337WK2, 00001028338WK0, 00001028339WK8, 00001028340WK6, 00001028341WK4, 00001028342WK2, 00001028343WK0, 00001028344WK8, 00001028345WK5, 00001028346WK3, 00001028347WK1, 00001028351WK3, 00001028352WK1, 00001028353WK9, 00001028354WK7,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDATWISTER PLUS 26mm Rotatable Retrieval Device (Box of 10 configuration) Intended to be used in flexibleJanuary 22, 2014
- FDAAnspach Carbide Cutting Burrs: 2MM Carbide Fluted Ball, Standard, 8 Flutes - part #QD8-2SB-C-8F, 4MM CarbideJanuary 22, 2014
Questions about this recall
Was GE Optima XR220amx and Optima XR200amx with Digital Upgrade Mobile Radiographic Systems. Model numbers 5555000-5, 5555000-6, and 5421698. recalled?
Yes. FDA published recall Z-0725-2014 on January 22, 2014. Reported hazard: There exists a software issue associated with patient selection from the worklist on the Optima XR200 with Digital Upgrade, and Optima XR220amx Mobile X-ray Systems. Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the mobile x-ray system are not properly displayed. If the user continues to attempt to select one of the exams w
What should I do if I have this product?
Voluntary: Firm initiated · Letter