GE Optima XR220amx and Optima XR200amx with Digital Upgrade Mobile Radiographic Systems. Model numbers 5555000-5, 5555000-6, and 5421698.

Published January 22, 2014 · Recall initiated: November 7, 2013

Summary

FDA records medical device Class II recall Z-0725-2014, published January 22, 2014 and initiated November 7, 2013. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 1027; Lot: Serial Numbers: 00000002796EN1, 00000002797EN9, 00000002798EN7, 00000002799EN5, 00000002800EN1, 00000002801EN9, 00000102726WK9, 00000102789WK3, 00000102937WK7, 00000130778WK3, 00000UA45526-8, 00001027058WK5, 00001027061WK9, 00001027146WK8, 00001027149WK2, 00001027150WK0, 00001027151WK8, 00001027153WK4, 00001027154WK2, 00001027167WK4, 00001027168WK2, 00001027169WK0, 00001027175WK7, 00001027193WK0, 00001027194WK8, 00001027195WK5, 00001027199WK7, 00001027200WK3, 00001027203WK7, 00001027204WK5, 00001027207WK8, 00001027214WK4, 00001027256WK5, 00001027259WK9, 00001027260WK7, 00001027262WK3, 00001027264WK9, 00001027265WK6, 00001027267WK2, 00001027268WK0, 00001027269WK8, 00001027276WK3, 00001027277WK1, 00001027278WK9, 00001027280WK5, 00001027293WK8, 00001027294WK6, 00001027295WK3, 00001027297WK9, 00001027298WK7, 00001027299WK5, 00001027300WK1, 00001027301WK9, 00001027302WK7, 00001027303WK5, 00001027314WK2, 00001027318WK3, 00001027322WK5, 00001027323WK3, 00001027324WK1, 00001027325WK8, 00001027326WK6, 00001027344WK9, 00001027346WK4, 00001027347WK2, 00001027349WK8, 00001027350WK6, 00001027351WK4, 00001027352WK2, 00001027353WK0, 00001027354WK8, 00001027356WK3, 00001027359WK7, 00001027362WK1, 00001027363WK9, 00001027364WK7, 00001027369WK6, 00001027371WK2, 00001027372WK0, 00001027373WK8, 00001027374WK6, 00001027376WK1, 00001027377WK9, 00001027380WK3, 00001027382WK9, 00001027383WK7, 00001027384WK5, 00001027386WK0, 00001027387WK8, 00001027388WK6, 00001027389WK4, 00001027390WK2, 00001027406WK6, 00001027407WK4, 00001027416WK5, 00001027417WK3, 00001027418WK1, 00001027421WK5, 00001027422WK3, 00001027423WK1, 00001027438WK9, 00001027445WK4, 00001027450WK4, 00001027451WK2, 00001027452WK0, 00001027464WK5, 00001027465WK2, 00001027466WK0, 00001027487WK6, 00001027488WK4, 00001027489WK2, 00001027490WK0, 00001027506WK3, 00001027507WK1, 00001027508WK9, 00001027539WK4, 00001027540WK2, 00001027541WK0, 00001027542WK8, 00001027543WK6, 00001027544WK4, 00001027551WK9, 00001027552WK7, 00001027553WK5, 00001027582WK4, 00001027583WK2, 00001027699WK6, 00001027700WK2, 00001027736WK6, 00001027737WK4, 00001027749WK9, 00001027762WK2, 00001027764WK8, 00001027765WK5, 00001027766WK3, 00001027768WK9, 00001027769WK7, 00001027772WK1, 00001027773WK9, 00001027774WK7, 00001027775WK4, 00001027778WK8, 00001027779WK6, 00001027781WK2, 00001027814WK1, 00001027819WK0, 00001027820WK8, 00001027821WK6, 00001027823WK2, 00001027824WK0, 00001027825WK7, 00001027826WK5, 00001027827WK3, 00001027828WK1, 00001027878WK6, 00001027880WK2, 00001027887WK7, 00001027888WK5, 00001027889WK3, 00001027890WK1, 00001027891WK9, 00001027892WK7, 00001027893WK5, 00001027894WK3, 00001027895WK0, 00001027896WK8, 00001027915WK6, 00001027932WK1, 00001027934WK7, 00001027935WK4, 00001027939WK6, 00001027992WK5, 00001027993WK3, 00001027994WK1, 00001027995WK8, 00001027996WK6, 00001028003WK0, 00001028021WK2, 00001028022WK0, 00001028023WK8, 00001028025WK3, 00001028035WK2, 00001028036WK0, 00001028042WK8, 00001028043WK6, 00001028044WK4, 00001028050WK1, 00001028051WK9, 00001028052WK7, 00001028053WK5, 00001028064WK2, 00001028065WK9, 00001028066WK7, 00001028077WK4, 00001028078WK2, 00001028079WK0, 00001028080WK8, 00001028091WK5, 00001028092WK3, 00001028093WK1, 00001028094WK9, 00001028119WK4, 00001028120WK2, 00001028205WK1, 00001028213WK5, 00001028214WK3, 00001028215WK0, 00001028216WK8, 00001028217WK6, 00001028230WK9, 00001028231WK7, 00001028232WK5, 00001028267WK1, 00001028269WK7, 00001028275WK4, 00001028276WK2, 00001028277WK0, 00001028279WK6, 00001028280WK4, 00001028281WK2, 00001028283WK8, 00001028284WK6, 00001028285WK3, 00001028286WK1, 00001028287WK9, 00001028288WK7, 00001028289WK5, 00001028290WK3, 00001028291WK1, 00001028292WK9, 00001028294WK5, 00001028331WK5, 00001028333WK1, 00001028334WK9, 00001028337WK2, 00001028338WK0, 00001028339WK8, 00001028340WK6, 00001028341WK4, 00001028342WK2, 00001028343WK0, 00001028344WK8, 00001028345WK5, 00001028346WK3, 00001028347WK1, 00001028351WK3, 00001028352WK1, 00001028353WK9, 00001028354WK7,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There exists a software issue associated with patient selection from the worklist on the Optima XR200 with Digital Upgrade, and Optima XR220amx Mobile X-ray Systems. Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the mobile x-ray system are not properly displayed. If the user continues to attempt to select one of the exams w

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Worldwide Distribution-USA (nationwide) including Washington D.C., and Guam and the countries of ALBANIA, AUSTRALIA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CHILE, CHINA , COLOMBIA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, IRAQ, IRELAND, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MEXICO, NEW ZEALAND, NIGERIA, NORWAY, POLAND, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, UNITED ARAB EMIRATES, and UNITED KINGDOM.
Lot
Serial Numbers: 00000002796EN1, 00000002797EN9, 00000002798EN7, 00000002799EN5, 00000002800EN1, 00000002801EN9, 00000102726WK9, 00000102789WK3, 00000102937WK7, 00000130778WK3, 00000UA45526-8, 00001027058WK5, 00001027061WK9, 00001027146WK8, 00001027149WK2, 00001027150WK0, 00001027151WK8, 00001027153WK4, 00001027154WK2, 00001027167WK4, 00001027168WK2, 00001027169WK0, 00001027175WK7, 00001027193WK0, 00001027194WK8, 00001027195WK5, 00001027199WK7, 00001027200WK3, 00001027203WK7, 00001027204WK5, 00001027207WK8, 00001027214WK4, 00001027256WK5, 00001027259WK9, 00001027260WK7, 00001027262WK3, 00001027264WK9, 00001027265WK6, 00001027267WK2, 00001027268WK0, 00001027269WK8, 00001027276WK3, 00001027277WK1, 00001027278WK9, 00001027280WK5, 00001027293WK8, 00001027294WK6, 00001027295WK3, 00001027297WK9, 00001027298WK7, 00001027299WK5, 00001027300WK1, 00001027301WK9, 00001027302WK7, 00001027303WK5, 00001027314WK2, 00001027318WK3, 00001027322WK5, 00001027323WK3, 00001027324WK1, 00001027325WK8, 00001027326WK6, 00001027344WK9, 00001027346WK4, 00001027347WK2, 00001027349WK8, 00001027350WK6, 00001027351WK4, 00001027352WK2, 00001027353WK0, 00001027354WK8, 00001027356WK3, 00001027359WK7, 00001027362WK1, 00001027363WK9, 00001027364WK7, 00001027369WK6, 00001027371WK2, 00001027372WK0, 00001027373WK8, 00001027374WK6, 00001027376WK1, 00001027377WK9, 00001027380WK3, 00001027382WK9, 00001027383WK7, 00001027384WK5, 00001027386WK0, 00001027387WK8, 00001027388WK6, 00001027389WK4, 00001027390WK2, 00001027406WK6, 00001027407WK4, 00001027416WK5, 00001027417WK3, 00001027418WK1, 00001027421WK5, 00001027422WK3, 00001027423WK1, 00001027438WK9, 00001027445WK4, 00001027450WK4, 00001027451WK2, 00001027452WK0, 00001027464WK5, 00001027465WK2, 00001027466WK0, 00001027487WK6, 00001027488WK4, 00001027489WK2, 00001027490WK0, 00001027506WK3, 00001027507WK1, 00001027508WK9, 00001027539WK4, 00001027540WK2, 00001027541WK0, 00001027542WK8, 00001027543WK6, 00001027544WK4, 00001027551WK9, 00001027552WK7, 00001027553WK5, 00001027582WK4, 00001027583WK2, 00001027699WK6, 00001027700WK2, 00001027736WK6, 00001027737WK4, 00001027749WK9, 00001027762WK2, 00001027764WK8, 00001027765WK5, 00001027766WK3, 00001027768WK9, 00001027769WK7, 00001027772WK1, 00001027773WK9, 00001027774WK7, 00001027775WK4, 00001027778WK8, 00001027779WK6, 00001027781WK2, 00001027814WK1, 00001027819WK0, 00001027820WK8, 00001027821WK6, 00001027823WK2, 00001027824WK0, 00001027825WK7, 00001027826WK5, 00001027827WK3, 00001027828WK1, 00001027878WK6, 00001027880WK2, 00001027887WK7, 00001027888WK5, 00001027889WK3, 00001027890WK1, 00001027891WK9, 00001027892WK7, 00001027893WK5, 00001027894WK3, 00001027895WK0, 00001027896WK8, 00001027915WK6, 00001027932WK1, 00001027934WK7, 00001027935WK4, 00001027939WK6, 00001027992WK5, 00001027993WK3, 00001027994WK1, 00001027995WK8, 00001027996WK6, 00001028003WK0, 00001028021WK2, 00001028022WK0, 00001028023WK8, 00001028025WK3, 00001028035WK2, 00001028036WK0, 00001028042WK8, 00001028043WK6, 00001028044WK4, 00001028050WK1, 00001028051WK9, 00001028052WK7, 00001028053WK5, 00001028064WK2, 00001028065WK9, 00001028066WK7, 00001028077WK4, 00001028078WK2, 00001028079WK0, 00001028080WK8, 00001028091WK5, 00001028092WK3, 00001028093WK1, 00001028094WK9, 00001028119WK4, 00001028120WK2, 00001028205WK1, 00001028213WK5, 00001028214WK3, 00001028215WK0, 00001028216WK8, 00001028217WK6, 00001028230WK9, 00001028231WK7, 00001028232WK5, 00001028267WK1, 00001028269WK7, 00001028275WK4, 00001028276WK2, 00001028277WK0, 00001028279WK6, 00001028280WK4, 00001028281WK2, 00001028283WK8, 00001028284WK6, 00001028285WK3, 00001028286WK1, 00001028287WK9, 00001028288WK7, 00001028289WK5, 00001028290WK3, 00001028291WK1, 00001028292WK9, 00001028294WK5, 00001028331WK5, 00001028333WK1, 00001028334WK9, 00001028337WK2, 00001028338WK0, 00001028339WK8, 00001028340WK6, 00001028341WK4, 00001028342WK2, 00001028343WK0, 00001028344WK8, 00001028345WK5, 00001028346WK3, 00001028347WK1, 00001028351WK3, 00001028352WK1, 00001028353WK9, 00001028354WK7,
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was GE Optima XR220amx and Optima XR200amx with Digital Upgrade Mobile Radiographic Systems. Model numbers 5555000-5, 5555000-6, and 5421698. recalled?

Yes. FDA published recall Z-0725-2014 on January 22, 2014. Reported hazard: There exists a software issue associated with patient selection from the worklist on the Optima XR200 with Digital Upgrade, and Optima XR220amx Mobile X-ray Systems. Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the mobile x-ray system are not properly displayed. If the user continues to attempt to select one of the exams w

What should I do if I have this product?

Voluntary: Firm initiated · Letter