Vital Diagnostics Envoy Calcium Reagent For In-Vitro Diagnostic Use. PN 55285 For the quantitative measurement of calcium in plasma and serum
Published January 22, 2014 · Recall initiated: August 29, 2013
Summary
FDA records medical device Class II recall Z-0732-2014, published January 22, 2014 and initiated August 29, 2013. Recalling firm: Vital Diagnostics, Inc.; Status: Terminated; Reported quantity: 395 units; Lot: All lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
High positive calcium bias on plasma sample versus results on serum
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Vital Diagnostics, Inc.
- Distributor
- Nationwide Distribution
- Lot
- All lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAConsult Diagnostics® Eon Calcium Reagent For In-Vitro Diagnostic Use PN: 1429 For the quantitative measurementJanuary 22, 2014
- FDACentricity PACS System versions 3.X and higher and 4.X and higher Product Usage: The Centricity PACSJanuary 22, 2014
- FDAPalacos® R+G Radioplaque Bone Cement 1 x 40 g Single (with Gentamicin), 40.8 g Methyl Acrylate Copolymer, 20January 22, 2014
- FDAOxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packsJanuary 22, 2014
- FDAAnspach Carbide Cutting Burrs & Drums: 2MM Carbide Fluted Ball, Extended part # S-2B-C and 7MM Carbide FlutedJanuary 22, 2014
- FDAAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System with stopcock and in-lineJanuary 22, 2014
Questions about this recall
Was Vital Diagnostics Envoy Calcium Reagent For In-Vitro Diagnostic Use. PN 55285 For the quantitative measurement of calcium in plasma and serum recalled?
Yes. FDA published recall Z-0732-2014 on January 22, 2014. Reported hazard: High positive calcium bias on plasma sample versus results on serum
What should I do if I have this product?
Voluntary: Firm initiated · Letter