Oxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packs containing 5 cartridges per pack, 50 discs per cartridge. Firm name on the label is Oxoid, Ltd., Basingstoke, United Kingdom. Used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing to assist in the determination of a suitable antibiotic agent for treatment purposes.
Published January 22, 2014 · Recall initiated: November 21, 2013
Summary
FDA records medical device Class II recall Z-0734-2014, published January 22, 2014 and initiated November 21, 2013. Recalling firm: Remel Inc; Status: Terminated; Reported quantity: 45/5/50 - discs/cartridges/packs; Lot: Lot number: 1223566 EXP: 8/31/2013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Individual discs in the lot may not be sufficiently impregnated with antibiotic, resulting in smaller than expected zone sizes.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Remel Inc
- Distributor
- Nationwide Distribution including the states of AL, AZ, CA, CO, FL, HI, IA, IL, KY, LA, MD, MI, MN, MO, MS, NC, NY, OH, SC, TN, TX, VA, WI, and WV
- Lot
- Lot number: 1223566 EXP: 8/31/2013
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThermo SCIENTIFIC, Sensititre STP6F, Gram Negative IVD ASTDecember 14, 2022
- FDAThio Med w/Dex, Hem, Vit K (5ml) 100/PK Product Usage; Recommended for use in qualitative procedures as aMay 29, 2019
- FDAThio Med w/Dex, Hemin, Vit K (5ml) 20/PK Product Usage: Recommended for use in qualitative procedures as aMay 29, 2019
- FDABlood Agar (TSA w/ 5% Sheep Blood) Plate, For In Vitro Diagnostic Use, Remel, Lenexa, KS 66215 Product Usage:May 3, 2017
- FDAConsult Diagnostics® Eon Calcium Reagent For In-Vitro Diagnostic Use PN: 1429 For the quantitative measurementJanuary 22, 2014
- FDAAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System with stopcock and in-lineJanuary 22, 2014
Questions about this recall
Was Oxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packs containing 5 cartridges per pack, 50 discs per cartridge. Firm name on the label is Oxoid, Ltd., Basingstoke, United Kingdom. Used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing to assist in the determination of a suitable antibiotic agent for treatment purposes. recalled?
Yes. FDA published recall Z-0734-2014 on January 22, 2014. Reported hazard: Individual discs in the lot may not be sufficiently impregnated with antibiotic, resulting in smaller than expected zone sizes.
What should I do if I have this product?
Voluntary: Firm initiated · Letter