Back to search
FDAFDA · Food and drugs·Recall number Z-0742-2024

Medex 700PSI STOPCOCK W/ROTATOR, List Number MX4331R

Published January 24, 2024 · Recall initiated: November 10, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Smiths Medical Asd Inc
Distributor
Worldwide distribution - US Nationwide and the countries of Canada, UAE.
Lot
Lot Number: 4355397 4380735 4387234 4404740
Risk classification
Class II
Area (FDA)
device
Country
United States